OdaLog
®
Gas Logger User Manual
P/N: 10-1000
Revision: F, Issued: October 2019
Page: 1
1 CERTIFICATION / CONFORMITY
1.1 Explosion-Protected Electrical Equipment - Certification &
Performance Conformity
The following Authorities have granted intrinsic safety approval to the OdaLog
®
Logger (Type L2).
Authority
Certificate Number
Code
EX TC
IECEx ITA 08.0010X
Ex ia IIC T4 (T
AMB
= -20
o
C to +50
o
C) Ga IP66/68
PRESAFE
PRESAFE 19 ATEX 09596X
II 1 G, Ex ia IIC T4 Ga IP66/68
1.2 Conditions of Safe Use
F
OR SAFETY REASONS THIS EQUIPMENT MUST BE OPERATED AND SERVICED BY QUALIFIED PERSONNEL ONLY
.
READ INSTRUCTION MANUAL
,
UNDERSTAND AND ADHERE TO MANUFACTURER
’
S LIVE MAINTENANCE
PROCEDURES BEFORE OPERATING OR SERVICING
.
T
HE BATTERY SHALL ONLY BE REMOVED OR REPLACED IN A NON
-
HAZARDOUS AREA
.
S
UBSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY
.
O
NLY USE WITH
O
DA
L
OG
®
APPROVED BATTERY TYPE
.
B
ATTERY MUST BE REPLACED WITH ONE OF THE FOLLOWING TYPES LISTED IN
T
ABLE
1.
T
O AVOID BUILD
-
UP OF ELECTROSTATIC CHARGE
,
THE
O
DA
L
OG
T
YPE
L2
SHALL BE USED OR PLACED WHERE
IT WILL BE PROTECTED FROM INADVERTENT RUBBING AND SHALL ONLY BE CLEANED OR WIPED WITH A DAMP
CLOTH WHEN USED IN EXPLOSIVE ATMOSPHERES CLASSIFIED AS
G
ROUP
IIC.
T
HIS CONDITION DOES NOT
APPLY FOR
G
ROUPS
IIA
AND
IIB.
Table 1 Approved Battery Types
Manufactures Part
Number Code
Type
Size
Saft LS26500
Lithium
C
Tadiran TL-5920/S
Lithium
C
1.3 Electromagnetic compatibility (EMC) – Compliance Statement
The OdaLog
®
Type L2 and accessories comply with the emissions and immunity requirements of:
1.4 FCC Statement
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part
15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference
when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate
radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause
harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause
harmful interference in which case the user will be required to correct the interference at his own expense.
1.5 Quality Statement
Thermo Fisher Scientific Australia Pty Ltd produces this product in accordance with our accredited AS/NZ ISO
9001, and ISO/IEC 80079-34 Quality System.
Standard Number
Standard Title
Compliance
EN 50270: 2015
Electrical Magnetic Compatibility – Electrical apparatus for the detection
and measurement of combustible gases, toxic gases or oxygen.
Fully