Aurora-2 User’s Manual
Appendix:
Collection Mode
In collection mode, the NSOM tip will be used to gather light rather than act as an illumi-
nation source. Therefore, the light exiting the tip fiber must be guided onto a suitable
detector. If your are using a detector with a rather large area, such as a PMT, that might
simply mean pointing the cleaved end of the fiber at the active area. However, if you are
using an APD, for example, some imaging system has to be set up, as the beam exiting the
fiber (NA 0.1) will simply overfill the active detector area.
The following example describes a collection mode set-up using a Hamamatsu photon-
counting PMT module, as shown below.
After cleaving the fiber at approximately 3mm, put the fiber into the guiding groove of the
holder and insert it carefully into the 12mm-long tube in front of the detector. Make sure
not to break the fiber by touching the detector window. Secure the fiber in the guiding
groove using the two magnets.
You are now ready to go into feedback and conduct your experiments.
Collection Mode Set-up with Hamamatsu Photon Counting PMT