background image

SPECIFICATIONS

LED FLASHING MODES

LCD INDICATORS

 

(TW-USB-1-LCD, TW-USB-2-LCD, TW-USB-2-LCD+ & TW-USB-TC-LCD ONLY)

Green double flash

Every 20 seconds: logger full, no alarm

Green single flash 

Every 30 seconds: delayed start

Every 20 seconds: low battery, logging, no alarm

Every 10 seonds: logging, no alarm

Red/Green single flash 

Every 20 seconds: logger full, last value equal to or 

exceeds the low or high alarm

Red double flash 

Every 10 seconds: logging, high alarm

Every 20 seconds: logging, high alarm, low battery

Red single flash 

Every 10 seconds: logging, low alarm

Every 20 seconds: logging, low alarm, low battery

No LEDs flash 

LEDs disabled during configuration to save battery 

power and increase logging life /logger stopped / 

battery empty

TW-USB-1-LCD, TW-USB-1-RCG,

TW-USB-TC & TW-USB-TC-LCD

Green double flash 

Delayed start 

Red double flash

High alarm

Orange double/single flash 

  

Memory full

Green single flash 

  Logging, no 

alarm

Red single flash 

      

Low alarm

Orange single flash every 60 sec       

 

Low battery

No LEDs flash 

Logger stopped or battery empty

TW-USB-1

TW-USB-2, TW-USB-2+, 

TW-USB-2-LCD & TW-USB-2-LCD+

Green triple flash

  

Logger full, no alarm on the channel

Red triple flash 

      

Logger full, alarm on the channel

Green double flash  

                 

Delayed start

Green single flash

  

Logging, no alarm

Red single flash 

  

Logging, low alarm on the channel

Red double flash

 

Logging, high alarm on the channel

No LEDs flash

 

  

Logger stopped or battery empty

Temperature scale symbols

Maximum/Minimum 
logged value symbol

LCD INDICATION

 

Display 

     Logger Status                  

Explanation

Delayed Start

Push to Start

Logging

Stopped

This is shown when the logger is set to 

start at a specific data and time*

This is shown when the logger is setup

for “Push to start” logging

This is shown when the logger is running 

in “LCD off” mode, and the button is 

pressed. The display clears again after

three seconds

If the logger has not been set to log and

the button is pressed, three dashes are 

displayed for three seconds

* If the logger is set to “LCD off” or “LCD on for 30 seconds” mode, then this will only be shown after the 

   button is pressed. Otherwise the display will remain blank.

DISPLAY AND STATUS FUNCTIONS

LCD INDICATION

TW-USB-1-LCD & TW-USB-2-LCD

TW-USB-TC-LCD

www.thermoworks.com

Part No

Function

Range

Accuracy (overall error)

Typ.                    Max.

Readings

Battery

Battery Life*

TW-USB-1

Temperature

–35 to 80°C (–31 to 176°F)

±1ºC (±2ºF)

16,382

3.6V ½AA

1 year

TW-USB-1-LCD

Temperature with LCD

–35 to 80°C (–31 to 176°F)

±0.5ºC (±1ºF)

±1.5ºC (±3ºF)

16,382

3.6V ½AA

1 year

TW-USB-1-PRO

Extended Temperature

–40 to 125°C (–40 to 257°F)

±0.2ºC (±0.4ºF)

32,510

3.6V 

AA

3 years

TW-USB-1-RCG

Temperature with rechargeable 

battery

–20 to 60°C (–4 to 140°F)

±1ºC (±2ºF)

32,510

Lithium Ion

1 month 

(rechargeable)

TW-USB-2

Temperature, Humidity & Dew Point

–35 to 80°C (–31 to 176°F)

0 to 100%RH

±0.5ºC (±1ºF)

±3%RH

±2ºC (±4ºF)

±6.0%RH

16,382

16,382

3.6V ½AA

1 year

TW-USB-2+

Increased Accuracy Temperature, 

Humidity & Dew Point

–35 to 80°C (–31 to 176°F)

0 to 100%RH

±0.3ºC (±0.6ºF)

±2.0%RH

±1.5ºC (±3ºF)

±4.0%RH

16,382

16,382

3.6V ½AA

1 year

TW-USB-2-LCD

Temperature, Humidity and Dew 

Point with LCD

–35 to 80°C (–31 to 176°F)

0 to 100%RH

±0.5ºC (±1ºF)

±3.0%RH

±2ºC (±4ºF)

±6.0%RH

16,382

16,382

3.6V ½AA

1 year

TW-USB-2-LCD+ Increased Accuracy Temperature, 

Humidity and Dew Point with LCD

–35 to 80°C (–31 to 176°F)

0 to 100%RH

±0.3ºC (±0.6ºF)

±2.0%RH

±1.5ºC (±3ºF)

±4.0%RH

16,382

16,382

3.6V ½AA

1 year

TW-USB-TC

Thermocouple (J, K and T-type)

–200 to 1350ºC (–328 to 2462ºF) (K-type)

–200 to 1190ºC (–328 to 2174ºF)  (J-type)

–200 to 390ºC (–328 to 734ºF) (T-type)

±1ºC (±2ºF)

±2ºC (±4ºF)

32,510

3.6V ½AA

6 months

TW-USB-TC-LCD

Thermocouple with LCD 

(J, K and T-type)

–200 to 1350ºC (–328 to 2462ºF) (K-type)

–200 to 1190ºC (–328 to 2174ºF)  (J-type)

–200 to 390ºC (–328 to 734ºF) (T-type)  

±1ºC (±2ºF)

±2ºC (±4ºF)

32,510

3.6V ½AA

6 months

*Depending on logging rate, ambient temperature, and use of alarm LED

ThermoWorks

ThermoWorks, Inc.

741 E. Utah Valley Dr.

American Fork, UT 84003

UNITED STATES

T (801) 756-7705

F (801) 756-8948

E orders@thermoworks.com

Temperature 

scale

Relative Humidity 

symbol*

Minimum logged 
value symbol

Maximum logged 
value symbol

Reading 

Negative symbol

*TW-USB-2-LCD only

Issue 3

P-12-018-02-b