7.2.13 Command FB
<id>FB<parameter><CR>
Quick start mode
Access:
Reading / writing
Description:
With the command “FB”, the quick start mode is set.
Parameter description:
0: quick start mode switched off
1: quick start mode switched on
Value range:
0...1
Initial value:
1
7.2.14 Command HI
<id>HI<parameter><CR>
Hardware ID
Access:
Reading
Description:
With the command “HI”, the 128-bit hardware identification
number (clear identification) is read.
Parameter description:
32 places in hexadecimal representation
Example
533331003846344D3330313034363038
Value range:
Initial value:
7.2.15 Command ID
<id>ID<parameter><CR>
Identification number
Access:
Reading / writing
Description:
This command sets the identification number (THIES
interpreter) and/or the slave address (MODBUS RTU
Interpreter). A response telegram is only sent if the ‘id’
contained in the command agrees with the one set in the
weather station. An exception to this is the generic ‘id’, to which
all weather stations respond (THIES Interpreter). After the ‘id’
has been changed the device responds immediately with the
new ‘id’.
Parameter description:
99
generic ‘id’ (THIES interpreter)
0
broadcast slave address (MODBUS RTU interpreter)
Value range:
0 to 99 (THIES interpreter)
1 to 247 (MODBUS RTU interpreter)
Initial value:
0 (THIES interpreter)
1 (MODBUS RTU interpreter)