5.3 Replacing the UVC Lamp
To assure sterilisation of the granulate, replace the UVC lamp after 8000 operating
hours. Replacement of the lamp is shown on the pictures below.
Before replacing the UVC lamp turn the machine off by switching the rocker switch to
“O” and disconnect the power plug from the electrical socket.
1. Open the cover and remove upper
securing bolt using a 4mm Hex
wrench (Allen key), and the lower
screw loosen (1-2 turns
counterclockwise)
2. Open the front flap
3.
Lamp by 90° (backwards)
4. Pull lamp
5. Insert new bulb and turn it 90°
(forward)
6. Close the front cover and screw
fastening screws and tighten
UVC-Lampe