20
rmP.U (Ramp Units):
Press the UP or DOWN ARROW
to toggle between ramp units (SEc, min or Hour).
Controller will default to “SEc.”
dwL.U (Dwell Units):
Press the UP or DOWN ARROW
to toggle between dwell units (SEc, min or Hour).
Controller will default to “SEc.”
Cyc.n (Cycle Number):
Press the UP or DOWN
ARROW to set the number of program cycles (1 to 999
or cont). The controller will default to “cont.”
SEG.n (Segment Number):
Press the UP or DOWN
ARROW to select the segment number (1-8).
tYPE (Type of Segment):
Press the UP or DOWN
ARROW to select the segment type [End (end of pro-
gram), rmP.r = ramp rate (ramp to a specified setpoint at
a set rate), rmp.t = ramp time (ramp to a specified tem-
perature in a set time), dwEll (to maintain a constant
temperature for a set time), StEP (climb instantaneously
from current to specified temperature). The controller will
default to “End.” Other parameters used with tYPE
include; tGt target setpoint), Rate (rate of temperature
increase) and dur (time to target setpoint or time to
dwell).
End.t (End Type):
End segment type: dwELL (dwell con-
tinuous), rSEt (reset) and S OP (End Segment Output
power level.
AL LiSt (Alarm List)
IdHi:
Deviation High Alarm. Factory preset to 50.
Atun LiSt (Autotune List)
tunE:
One-shot autotune enable. Resets the PID values
automatically.
drA:
Adaptive tune enable. Monitors the process variable
and resets the PID values if the process variable
exceeds the trigger level.
drA.t:
Adaptive tune trigger level in display units. Range
= 1 9999.
8 S
EGMENT
P
ROGRAMMABLE
M
ODEL W
/OTP