© 2011 Thorlabs
38
Thorlabs Beam 4.0
The X profile displays a single pixel row taken from the received camera image,
whereas the Y profile shows a single pixel column. The column and row is defined by
the position of cross hair within the
graph.
The yellow graph shows the measured profile, while the red curve shows the
approximated Gaussian fit function. If "Autoscale to Peak" is enabled and in 2D
projection the cross hair is fixed to peak, the measured curve shows relative
intensities from 0 to 100%, where 100% denotes the maximum value in the selected
row / column. The amplitude of the Gaussian fit curve may be lower or even higher
than the peak intensity of the measured curve.
The selected clip level (default 13.5%) is displayed, if the "Auto Scale to Peak"
function is enabled (button
).
The horizontal scale is displayed in pixels or µm and its range refers to the selected
. The unit of the scale can be changed with the
dialog.
Toolbar
Icon
Associated Action
Opens a dialog box to specify the properties of the saved screenshots /
diagrams.
Opens a dialog box to save measurement data to XLS or CSV file.
Toggle button to display grid in the diagram. Default: grid is shown.
Zoom Home button
Toggle button to show or hide the cursor.
Toggle button to show or hide the Gaussian curve fit. Default: curve fit is
shown.
Zoom Mode
33
53
59
Summary of Contents for BC106-UV
Page 1: ...2011 Operation Manual Thorlabs Beam Beam Analyzing Software BC106 VIS BC106 UV ...
Page 2: ...Version Date 4 0 11 04 2011 2011Thorlabs 2011Thorlabs ...
Page 76: ... 2011Thorlabs 74 Thorlabs Beam 4 0 Example ...
Page 157: ... 2011Thorlabs 155 Appendix 7 4 4 Drawings 7 4 4 1 BC106 UV ...
Page 158: ... 2011Thorlabs 156 Thorlabs Beam 4 0 7 4 4 2 BC106 VIS ...
Page 159: ... 2011Thorlabs 157 Appendix 7 4 4 3 BC1M2 Mounting Adapter ...
Page 160: ... 2011Thorlabs 158 Thorlabs Beam 4 0 7 4 4 4 Translation Stage VT 80 ...
Page 167: ......