3 Operating Instruction
15
The display shows the injection current "I
LD
".
With the adjust knob (F8) the laser current can be set continuously between 0 mA and the
selected current limit "I
LIM
".
ð
If the injection current "I
LD
" reaches the set current limit "I
LIM
" during operation, the LED
"LIMIT" (F5) lights up, a short alert beep is heard, and the laser current is limited to the value
of the current limit "I
LIM
". In this case, noise a nd ripple do not longer correspond to the
specifications for normal operation. However exceeding the set current limit "I
LIM
" is
impossible.
ð
If the connection to the laser diode is interrupted during operation, the laser diode current is
switched off automatically. LED "LASER ON" (F7) extinguishes, LED "OPEN" (F4) lights up
and a short beep is heard.
ð
If the output is switched on while the interlock is closed and there is no laser diode
connected, at first the LED "LASER ON" (F7) lights up and the output is switched on. Then
the LDC2xxC recognizes the missing laser diode and switches the output off. The LED
"LASER ON" (F7) extinguishes and the LED "OPEN" (F4) lights up. Connect the laser and
press "LASER ON" (F8) - the LED "OPEN" (F4) extinguishes and the output switches on.
ð
If a photodiode is connected, the display can be set to show the photodiode current "I
PD
" or
the optical power "P
LD
" by pressing the key (F17) or (F18).
The laser current can be
via the connector "MOD IN" (R2).
3.4 Constant Power Mode (CP)
If the laser includes a photodiode (also known as monitor diode), the laser can be operated
also in constant power mode. In contrast to the constant current mode, here the photodiode
current is kept constant by controlling the laser injection current.
·
Start operating the LDC2xxC in constant current mode (refer to chapter 3.4)
·
Select the display "I
PD
" with the key (F17) or (F18).
·
Check whether an appropriate photodiode current "I
PD
" is available. A photodiode current of
at least the minimum control current must be available for a stable operation of the power
control:
5 µA for LDC200CV,
25 µA for LDC201CU, LDC202C, LDC205C,
50 µA for LDC210C, LDC220C, LDC240C.
Note
If the sign of the photodiode current "I
PD
" is negative, the photodiode must be reversed.
·
Switch off the laser current by pressing the key "LASER ON" (F8) .
·
Select constant power mode (“P”) by pressing the key “CONST” (F24) until LED "P" (F21)
lights up.
·
Set the adjust knob (F9) completely counter clockwise.
·
Switch on the output by pressing "LASER ON" (F8). The LED "LASER ON" (F7) lights up,
the output is activated and the current slowly increases (about 1 s) to the set value.
·
The photodiode current "I
PD
" can be set using (F9) and with that the optical power of the
laser diode increases until the laser current "I
LD
" reaches the selected current limit "I
LIM
".
ð
If the desired photodiode current "I
PD
" cannot be set with a sufficient accuracy (F9), the
range of the knob (F9) can be fitted to the current range of the connected photodiode using
the potentiometer "PD RANGE" (F25) - see below.
ð
If no photodiode is connected or the polarity of the photodiode is set wrong, the laser current
17