LFE1220-US, -EU, and -UK Laminar Flow Enclosures
Chapter 3: Installation
Rev A August 2020
Page 9
2. The construction rails are assembled by sliding the head of low profile screws into the T-slots of the construction
rail. To enable the assembly, the spacing between the end of the construction rail and the underside of the screw
head must be set to a specific distance. For convenience, all assembly screws have been pre-fitted, but the exact
spacing must be set before assembly. An assembly tool (AO) has been provided for this purpose.
To set the spacing
: loosen the screws from their pre-installed position using the 4mm T-handle hex driver (AL),
insert the assembly tool (AO) under the screw heads, tighten the screws against the assembly tool, then back off
1/4 turn to release the assembly tool.
For this step
: set the spacing of the low profile screws in both ends of the Longitudinal Supports:
Figure 2 Using the Assembly Tool
3. Using the 4mm T-handle hex driver (AL) supplied, loosen the 3 screws holding the flange to the Lateral Filter
Support – just enough for the flange to slide along the slot (<1 turn). Slide the flange to one side.
Figure 3 Sliding the Flange