background image

E-M-HP21-V1_21 

Rotronic AG 
Bassersdorf, Switzerland 

Document code 

Unit 

HygroPalm HP21 hand-held indicator: 
User Guide 

Instruction Manual 

Document Type 

Page 

3

 of 14 

Document title 

 
 

© 2009-2010; Rotronic AG                                                                                                                              E-M-HP21-V1_21

 

Applicability: 

 
This manual is valid for the HP21 with firmware version 1.x, where 1.x can be 1.0, 1.1 etc. Changes to the last 
digit of the version number reflect minor firmware changes that do not affect the manner in which the 
instrument should be operated.  

1  

Overview 

The HP21 is a highly accurate hand-held indicator that displays relative humidity, temperature and the dew or 
frost point. The HP21 is perfect for spot checking HVAC installations, manufacturing and storage areas, and 
generally for all ambient air measurement within the temperature range of -10 to 60°C (14 to 140 °F). The 
HP21 operates with a regular 9V alkaline battery. 
 
Based on the ROTRONIC AirChip 3000 digital technology the HP21 offers the following user functions: 
 

  User configurable settings 
  Calculation of the dew or frost point 
  Humidity temperature calibration and adjustment 
  Simulator mode 
  Automatic humidity sensor test and drift compensation 
  Sensor failure mode 
  Data recording 

 
The ability for the user to easily update the AirChip 3000 firmware means that the HP21 can be kept up-to-
date regarding any future functionality improvement.  

2  

General description 

2.1    Power supply 

The HP21 operates with a regular 9V alkaline battery 

2.2    Measured parameters 

The HP21 measures relative humidity with a ROTRONIC Hygromer

®

 IN1 capacitive sensor and temperature 

with a Pt100 RTD. 

2.3    Calculated parameters 

Using the ROTRONIC HW4 software, the HP21 can be configured by the user to calculate either the dew 
point or the frost point. 

2.4    Sensor protection (dust filter) 

The HP21 is supplied with a Polyethylene filter to protect the sensors against dust particles and high air 
velocity. 

Summary of Contents for HygroPalm HP21

Page 1: ...sersdorf Switzerland Document code Unit HygroPalm HP21 hand held indicator User Guide Instruction Manual Document Type Page 1 of 14 Document title 2009 2010 Rotronic AG E M HP21 V1_21 HygroPalm HP21 H...

Page 2: ...tions 4 3 1 Function overview 4 3 2 Factory default settings 6 4 Operation 7 4 1 Display 7 4 2 Keypad 7 4 3 Internal menu 8 4 4 Changing the parameters shown on the display 8 4 5 Changing the temperat...

Page 3: ...V alkaline battery Based on the ROTRONIC AirChip 3000 digital technology the HP21 offers the following user functions User configurable settings Calculation of the dew or frost point Humidity temperat...

Page 4: ...t internal menu to do for example a humidity and temperature adjustment of the HP21 3 User configurable settings and functions The HP21 can be used just as any conventional humidity and temperature in...

Page 5: ...hen the device is powered off the recording session is paused but not ended As long as the recording session has not been ended the device automatically resumes recording data when powered up again o...

Page 6: ...esolution 1 decimal Display backlight K On Key Press Displayed parameters K RH and temperature Trend indicator display Enabled Communication protocol RO ASCII RS 485 address 0 Device name Instrument m...

Page 7: ...idity or dew frost point and the bottom line corresponds to temperature The display can be configured to show a trend indicator on each line increasing value decreasing value In the event of an alarm...

Page 8: ...nt adjustment only offset Notes o Record both the recording mode start stop and the log interval cannot be changed from the menu and are as configured with the ROTRONIC HW4 software o SensorTest Off m...

Page 9: ...ensation may occur on the sensors when the probe is colder than the environment As long as the humidity temperature limits of the humidity sensor are not exceeded condensation does not alter the calib...

Page 10: ...to select either H Adjust or T Adjust and press the ENTER key Make sure that the text line beginning with RefValue is highlighted and press the ENTER key Use the UP or DOWN key to change the reference...

Page 11: ...easily If this happens correct the alignment by holding the sensor very gently with a pair of small flat nosed pliers 6 Firmware updates Firmware updates will be available on the ROTRONIC website for...

Page 12: ...IEC 751 1 3 class B Measuring range 10 60 C Measurement accuracy at 23 C 0 2 C Repeatability 0 05 C Long term stability 0 1 C year Sensor time constant Typical 4 sec 63 of a step change 1m sec air fl...

Page 13: ...nsing Operating limits at electronics 10 60 C limited by LC display 0 100 RH non condensing Temperature limits at probe Same as electronics Maximum humidity at probe Same as electronics Maximum air ve...

Page 14: ...Main HW4 software version 3 General instructions and functions common to all devices E M HW4v3 F2 004 HW4 software version 3 HP21 hand held indicator Device configuration and AirChip 3000 functions E...