MyoTrac Infiniti
™
User Guide
4
Type BF Equipment
Internally powered equipment
Continuous operation
Read Instruction Manual
The pins of the connectors identified with the ESD warning symbol should not be touched
unless ESB precautionary procedures are used.
CAUTION
US Federal Law restricts this device to sale by, or on order of, a physician or any other
practitioner licensed by the law of the state in which he or she practices to use or order the
use of this device.
WARNING
Do not operate Active Sensors within 10 feet of an operating cellular phone, similar radio
transmitting device, other powerful radio interference producing sources such as arc
welders, radio thermal treatment equipment, x-ray machines, or any other equipment that
produces electrical sparks. Portable and mobile RF communication equipment can affect
this equipment.
With the MyoTrac Infiniti Encoder SA9800 use only with supplied power supply. GlobTek
Part Number WR92B2500LF9P-Y-MED (WR95/WR93/WR97) or GS889
The PC used with MyoTrac Infiniti must be placed outside the patient/client environment
(more than 3 meters or 10 feet) or the PC must comply with EN60601-1 (system safety).
After use, the Batteries or the Battery pack must be disposed of in accordance with local,
state and federal regulations and laws.
After use, the Disposable Electrodes may be a biohazard. Handle, and when applicable,
dispose of these materials in accordance with accepted medical practice and any applicable
local, state and federal laws and regulations.
Reusable electrodes present a potential risk of cross-infection especially when used on
abraded skin, unless they are restricted to a single patient or sterilized between patients. If
sterilizing electrodes, employ only gas sterilization.
Radiated radio frequency electromagnetic fields can cause performance degradation in the
MyoScan EMG sensor. In the worst case, an RF field strength of 22mV/M can cause an
increase of 1
V in the signal reading from a MyoScan sensor. Be sure to keep in mind that
a very relaxed muscle should provide an EMG reading of approximately 1-3
V.
This device is capable of generating current densities exceeding 2mA r.m.s./cm
² this may
require special attention of the operator.
Do not exceed 0.1watts/cm
² with recommended electrodes, assuming a load of 500
.
1.
1’’ Round = Max 69mA
2.
2’’ x 2’’ Square = Max 100mA
3.
3’’ x 4’’ Rectangle = Max 100mA
4.
St Cloud = Max 100mA
5.
Femelex = Max 100mA
Avoid accidental contact between connected but unused applied parts and other conductive
parts including those connected to protective earth.
Explosion Hazard; Do not use in the presence of a flammable anesthetic mixture with air, or
with Oxygen or Nitrous Oxide.
Not to be immersed in water.
Take care in arranging patient and sensor cables to avoid risk of patient entanglement or
strangulation.
The operator is responsible for ensuring the safety of any devices controlled or triggered by
Infiniti equipment or software, or by any software or hardware receiving data from Infiniti
equipment. Infiniti equipment must not be configured or connected in such a way that
failure in its data acquisition, processing or control functions can trigger patient feedback
stimulus that poses an unacceptable level of risk.
Use of any equipment in a biofeedback or stimulation context should be immediately