MyoTrac Infiniti
™
User Guide
40
To run an EMG-Stim Script select:
New Script Session
from the main menu.
Select EMG-Stim
Select Body area from illustration for area specific scripts.
Select the threshold mode
o
EMG-Triggered Stim (threshold manually set)
o
EMG-Triggered Stim (threshold automatically set)
o
Alternating Stim and EMG
Select Script by using the up, down and forward arrows.
A number of pre-defined Scripts are
available or you can design your own using the custom design Wizard, covered later in the
chapter.
Threshold Setting: Confirm or change the threshold that will be used to determine if the
stimulation should be triggered.
Set Stim Intensity:
The intensity that is set here will be delivered if the correct criteria are met
in the recording phase. Use the up and down arrows to set the intensity. When this is set
move forward to the recording screen.
Start the EMG Triggered Stim session:
Start the Script driven session by tapping the record
button. To stop the session tap this button again, or, during the Stim delivery, press the stop
button.
Note:
EMG triggered STIM is not available for incontinence application
.