Recommended separation distances between
portable and mobile RF communications equipment and the MyoTrac Infiniti
The MyoTrac Infiniti is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the MyoTrac Infiniti can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the
MyoTrac Infiniti as recommended below, according to the maximum output power of the communications
equipment.
Rated maximum output
power of transmitter
W
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2,5 GHz
P
d
2
.
1
P
d
2
.
1
P
d
3
.
2
0,01
0.12
0.12
0.23
0,1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in
metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where
P
is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Guidance and manufacturer’s declaration – electromagnetic emissions
The MyoTrac Infiniti is intended for use in the electromagnetic environment specified below. The customer or the user of
the MyoTrac Infiniti should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment
– guidance
RF emissions
CISPR 11
Group 1
The MyoTrac Infiniti uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The MyoTrac Infiniti is suitable for use in all establishments,
including domestic establishments and those directly connected to
the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable