background image

 

16

APPENDICES 

APPENDIX A. THE OSCILLOSCOPE DISPLAY

 

Single or dual cable connection 

The  DSA524  produces  a  multiplexed  output  which  enables  two 
traces  plus  text  to  be  displayed  using  only  one  channel  of  the 
display  oscilloscope.  The  multiplexed  output  also  incorporates  a 
trigger  signal.  Consequently  the  DSA  can  be  connected  to  the 
oscilloscope via just one cable with triggering being sourced from 
the input channel. 

The  disadvantage  of  single  cable  connection  is  that  the  trigger 
signal  is  visible  at  the  left  hand  edge  of  the  screen  as  a  very 
narrow vertical pulse followed by a short bright line. Connecting a 
second  cable  from  the  "Trig  out"  socket  to  the  external  trigger 
input of the oscilloscope and setting the scope to external trigger 
removes the visible trigger signal from the screen. 

Triggering 

Because  the  DSA524  produces  a  multiplexed  output  waveform, 
trigger  adjustment  is  very  important  particularly  when  text  is 
being  displayed.  If  a  satisfactory  initial  set-up  display  cannot  be 
obtained check the following points: 

The  'scope  must  be  set  to  single  channel  operation.  The  input 
coupling  must  be  DC  and  the  sensitivity  must  be  set  to 
100mV/div.  The  sweep  hold-off  and  trigger  delay  (if  present) 
must  be  set  to  minimum.  The  sweep  speed  must  be  set  to 
50us/div. The trigger source must be from the input channel (not 
vertical  mode).  The  trigger  slope  and  level  must  both  be 
negative. 

Adjust  the  trigger  level.  If  a  satisfactory  display  can  still  not  be 
obtained try setting the sweep speed to 20us/div and then slow-
ing it down to 50us/div using the sweep variable control. 

When using a single cable connection, some oscilloscopes (those 
having  a  poor  dynamic  range  on  their  trigger  circuit)  may  give 
unstable triggering when full screen height waveforms are  being 
displayed.  Should  this  occur,  first  try  careful  adjustment  of  the 
trigger  level  control  but  if  the  trouble  persists  change  to  a  two 
cable connection. 

Vertical Size 

The  amplitude  of  the  vertical  output  is  factory  adjusted  to  an 
accuracy of + 1 %. The two horizontal lines of the set-up display 
are intended to be exactly 6 divisions apart. The output level can 
be  adjusted  to  compensate  for  inaccuracies  of  the  oscilloscope 
using  the  rear  panel  preset  adjustment  marked  "Set  Height".  If 
the  linearity  of  the  oscilloscope  is  poor,  however,  it  may  not  be 
possible  to  set  the  two  lines  at  exactly  ±3  divisions  from  the 
arrow  tip.  If  so  ignore  the  vertical  position  of  the  arrow  tip  and 
set  the  two  lines  to  be  at  ±3  divisions  from  the  graticule  centre 
line. 

The  horizontal  timing  of  the  DSA524  is  crystal  controlled  to  an 
accuracy  of  0.01%  (except  for  "repeat"  mode  operation)  and 
cannot be adjusted. 

APPENDIX B. ALIASING 

One  problem  that  occurs  with  digital  storage  oscilloscopes  that 
does not occur with real-time oscilloscopes is aliasing. 

Sampling  theory  dictates  that  any  periodic  waveform  must  be 
sampled at more than twice the highest frequency component to 
avoid aliasing. Aliasing causes the apparent waveform frequency 
to be much lower than the real frequency. Thus if the timebase is 
set too low an invalid display will result (see Fig a). 

To avoid aliasing the timebase must be set as follows: 

Minimum  time/div  =  50  divided  by  the  maximum  cyclic  signal 
frequency. 

The  difficulties  occur  when  the  signal  frequency  is  unknown.  In 
this  case  it  is  safest  to  set  a  fast  timebase  speed  initially  and 
work down. 

A  second  form  of  aliasing  that  can  occur  is  visual  aliasing.  This 
occurs  when  the  display  is  made  up  of  bright  dots  and  the 

number of samples per waveform period is low (though greater 
than two). The eye can be deceived into incorrectly linking the 
dots to form a slower period waveform (see Fig b). The dot join 
feature  of  the  DSA524  greatly  reduces  the  chances  of  visual 
aliasing occurring. 

 

 

Fig a. True aliasing - 
waveform sampled at 
less than twice the 
signal frequency (in 
this case 
approximately at the 
signal frequency). 

 

 

 

 

 

Fig b Visual aliasing -
vertical separation 
between samples 
large in comparison 
to horizontal separa-
tion, leading to incor-
rect interpretation by 
the eye. 

 

 

If the input signal is repetitive and the trigger repetition rate is 
between 50Hz and 5MHz the AUTOSET function of the DSA524 
can  be  used.  Autoset  resets  the  timebase  to  a  speed  suited  to 
the  trigger  rate  and  well  above  the  point  at  which  aliasing 
effects can occur. 

APPENDIX  C.  SINGLE-SHOT  BANDWIDTH  AND  INTER-
POLATION 

Probably the most important use of a digital storage instrument 
is  for  the  capture  of  single-shot  waveforms  (i.e.  waveforms 
which  occur  only  once).  The  user  will  need  to  understand  the 
bandwidth restrictions which apply to this mode of operation. 

Nyquist  sampling  theory  dictates  that  a  waveform  can  be  per-
fectly  re-constructed  if  sampled  at  more  than  twice  the  fre-
quency  of  the  highest  frequency  component.  Thus  a  9.9  MHz 
sinewave could be reconstructed if sampled at 20MHz. Unfortu-
nately this assumes that sampling can take place for an infinite 
period  of  time.  Clearly  this  is  not  possible  with  a  single-shot 
waveform. 

In  theory  a  single-shot  waveform  can  only  be  perfectly  recon-
structed  using  an  infinite  sampling  frequency.  This  is  because 
the  waveform  has  discontinuities  at  the  start  and  end  which 
give  rise  to  infinite  frequency  elements.  Given  a  known 
maximum  sampling  frequency  we  need  to  know  how  good  a 
representation of the original waveform can be achieved. 

Because there can be no certainty of the existence of any cyclic 
elements  in  a  single-shot  waveform,  sampling  theory  must  be 
abandoned.  Instead,  a  subjective  principle  must  be  used  to 
decide  what  constitutes  an  acceptable  representation  of  a 
waveform.  This  will  depend  on  the  information  that  the  user 
needs to obtain from the waveform. 

Consider  the  waveform  of  Fig  c  which  consists  of  a  rapid 
change of level, followed by an exponential decay of sinusoidal 
oscillations.  If  the  user  requires  only  to  know  the  settled 
amplitudes  before  and  after  the  change  of  level  the  sampling 
frequency  requirements  are  modest  since  a  representation 
which  merely  indicates  the  disturbance  and  the  settled  values 
will suffice. 

Summary of Contents for DSA524

Page 1: ...Thurlby DSA524 digital storage adaptor OPERATING MANUAL AND TUTORIAL V2 0 ...

Page 2: ...anual applies to units with a firmware level of 2 71 and above IT IS VERY STRONGLY RECOMMENDED THAT NEW USERS START BY FOLLOWING THE TUTORIAL INDEX Title Section Page REFERENCE SECTION Connecting an Oscilloscope R1 2 Setting up the Display R2 2 Front Panel Controls R3 2 Escape and Reset R4 2 Controlling Digitising R5 2 Timebase Setting R6 2 CH1 and CH2 R7 3 Digitising Memories R8 3 Trigger R9 3 Tr...

Page 3: ... NN the key will enter the number printed alongside it Note that when SHIFT or a function ending in NN is pressed the lamp marked ENTER will flash to indicate that one or more further key presses is required Any function that causes the ENTER lamp to flash can be cancelled by pressing that function key again Note that some of the functions including SET SCOPE automatically set the RUN HOLD functio...

Page 4: ... TRIGGER RATE ABOVE 1MHz and the DSA will beep to indicate that the timebase speed may need to be set faster R7 CH1 AND CH2 The two input channels are identical The input sensitivity is variable between 10V and 2mV per division The maximum bandwidth is 35MHz down to 20mV div 20MHz at 10mV div 10MHz at 5mV div and 5MHz at 2mV div The volts per division keys autorepeat if held The input coupling can...

Page 5: ... is automatically increased or decreased to maintain a constant time delay when the timebase speed is changed within a limit of 9 999 divisions maximum PRE there is no equivalent to pre trigger on a real time oscilloscope Pre trigger allows the part of the waveform which occurred before the trigger to be stored and observed The maximum pre trigger delay is 40 divisions Pressing either key shows th...

Page 6: ...unaffected R12 INDEXED WAVEFORM MEMORIES The DSA524 has sixteen 1K word memories for temporary or permanent storage of waveforms Each memory is identified by a two digit number from 01 through to 16 The memory is sup ported by trickle charged Ni Cad batteries which will remain charged for several months should the unit be left unpowered To store a waveform in an indexed memory obtain the required ...

Page 7: ...not captured the first section of the waveform captured starts 350 nanoseconds after the trigger event b Display Update Rate Repeat mode involves complex manipulation of data which can take several seconds per digitisation The display update rate varies between about one per second for a single channel at 2usecs div to one every 10 seconds for both channels at 50nsecs div The Rate key is disabled ...

Page 8: ...hen overflows could occur on the multiplied waveform The displayed waveform is limited to 4 25 divisions and conse quently 3 divs multiplied by 2 5 divs will be displayed as 4 25 divs instead of 7 5 divs The AxB function cannot operate while digitising is taking place Pressing the AxB key will set RUN HOLD to HOLD The function multiplies the traces exactly as they appear at the moment that the AxB...

Page 9: ... show baud rate 9600 for example Use the reference cursor keys keys 2 and 3 to increase or decrease the rate then press ESCAPE If the DSA is not connected to an oscilloscope the baud rate may be set directly using the FUNCTION NN key see Section R27 Note that as well as the options of 300 through to 38 400 there is a further option of IEEE This option is exclusively for use when the optional IEEE ...

Page 10: ...g SHIFT followed by RESET and with the signal generator being set to SINE 20kHz 5V pk pk INDEX TO THE TUTORIAL Page Equipment Required 10 Setting up the Equipment 10 First Steps 10 What to Expect 10 The Display Oscilloscope 10 Front Panel Controls 10 Getting out of Trouble 10 Part One The Basic Functions Controlling Digitising 10 Controlling the Input Channels 11 Controlling the Traces 11 Setting ...

Page 11: ...the display is made up of discrete steps waveforms do not look quite the same as they do with an analog oscilloscope The minimum level change between samples is one thirtieth of a division and even where the input signal level is constant the digitised level may vary slightly This variation is caused by noise on the signal noise in the input amplifiers or digitising noise in the ADC typically bit ...

Page 12: ...es on the screen and provide 4 25 divisions of adjustment Press the key marked HOME VAR to enable variable posi tion it is enabled when the lamp is off Try varying the position of the traces note that whereas OFFSET only affects the waveform when digitising is taking place PO SITION affects it at all times whether RUN HOLD is at RUN or HOLD Press the key again to home the trace position 2 The GAIN...

Page 13: ...els are on the ALT lamp illuminates for speeds of 10us div and faster and that the CHOP lamp illuminates for speeds of 20us div and slower CHOP indicates that both input channels are digitised together on opposite cycles of the sampling clock while ALT indicates that the input channels are digitised alternately CH2 followed by CH1 Thus if digitising is un triggered or if the input signals do not h...

Page 14: ... digitising memory captures 4096 words 40 96 divisions per channel The SEARCH key enables the whole of the digitising memory to be accessed It also provides digital magnification by ten using linear interpolation 1 Press SEARCH once the COMP compress lamp will illu minate and a filled bar will appear at the top of the screen for 3 seconds the filled bar represents the 4K of the memory Note that th...

Page 15: ... DSA reset and the generator set to SINE 20kHz 5V pk pk The DSA524 can average any number of readings from 2 up to 255 enabling large improvements to be made to the signal to noise ratio of noisy signals The average number is initially set to 16 1 Turn off Trace B and note the waveform amplitude on Trace A Then remove the trigger signal by setting the source to EXT The waveform should now be unsyn...

Page 16: ...volatile memory and replay them on demand Up to 50 settings can be stored Each front panel setting includes the state of every key and rotary control and the source of each trace waveform Thus if a setting is learned which includes a trace recalled from a wave form store that same store will be recalled again when the setting is re played Thus reference waveforms can be recalled automatically with...

Page 17: ...avoid aliasing Aliasing causes the apparent waveform frequency to be much lower than the real frequency Thus if the timebase is set too low an invalid display will result see Fig a To avoid aliasing the timebase must be set as follows Minimum time div 50 divided by the maximum cyclic signal frequency The difficulties occur when the signal frequency is unknown In this case it is safest to set a fas...

Page 18: ...isplay are plot ted using a level of 100mV div for both Y and X at a rate selec table as 1 2 5 or 10 seconds per division The pen lift signal is normally low to lift but can be inverted To set the plot rate press SHIFT key 8 followed by SET PLOT key 5 The oscilloscope display will show the present plot rate Press one of the REF CURS keys key 2 or 3 to select a new value then press ESCAPE If the DS...

Page 19: ...to have a buffer large enough to accept the data in one stream hand shaking will almost certainly be needed There are two basic types of handshaking hardware handshak ing and X ON X OFF handshaking Both types are provided on the DSA Hardware handshaking uses the RTS and CTS lines of the DSA When the RTS line is set to a 0 a positive voltage level the DSA is ready to receive data when it is set to ...

Page 20: ...trace starting at word N MAG N where N is a decimal number between 0 and 900 This sets the SEARCH function into Magnify mode with the trace starting at word N SROFF This turns any of the SEARCH modes off NORM SLOW or FAST This sets the display update rate ION or IOFF This sets sine interpolation on or off AON or AOFF This sets averaging on or off Example TMB 10uS AON SLOW CR For TRA or TRB the sec...

Page 21: ...ee previous section for an explanation of this command To load the data send the primary command MEM followed by a comma followed by one of the secondary commands listed below followed by a carriage return The secondary commands are as follows ACQ1 or AQU2 This loads the digitising memory of CH1 or CH2 4096 words must be sent TRA or TRB This loads the trace memory of Trace A or Trace B 1024 words ...

Page 22: ...n OSCILLOSCOPE DISPLAY SYSTEM Trace A Displays 1024 words from either CH1 or any indexed waveform storage memory Trace B Displays 1024 words from either CH2 or any indexed waveform storage memory Line Type Selectable as individual levels dots or joined to form a smooth line Update Rate Selectable between 50 per second and Run Mode one every 3 seconds continuous in roll mode OSCILLOSCOPE DISPLAY VE...

Page 23: ... dismantle it S4 FUSE REPLACEMENT The AC power fuse is mounted on the back panel within the AC line input socket housing To replace the fuse pull the clip in tray outwards The active fuse is at the back of the tray the front space is for a spare fuse The fuse must be replaced with a 500mA anti surge 20mm cartridge fuse A ceramic bodied HRC type should be used whenever possible S5 MAINTENANCE AND R...

Page 24: ...Thurlby Electronics Ltd New Road Burrel Road St Ives Huntingdon Cambs England Telephone 0480 63570 Fax 0480 64832 Telex 32475 ...

Reviews: