Synergy Controller Technical Manual, Revision H
Page 189
The Pressure display on Channel 2 is in scientific notation as shown below. Note that the Pressure Set Point
has no effect on Thermal Vacuum chambers since pressure is not controlled.
Pressure is recorded in the log file in Torr in scientific notation.
To verify setup and calibration verify the values in the following table.
Output in Torr is 10^((Vhires*2)-n)
n, Altitude Setting
V hi-res Volts DC
Display
12
5.0
1.0e-2
11
5.0
1.0e-1
10
5.0
1.0e-0
12
0.0
1.0e-12
11
0.0
1.0e-11
10
0.0
1.0e-10
12
2.1
1.6e-8
11
2.1
1.6e-7
10
2.1
1.6e-6
22.2 Altitude Chamber Setup
Summary of Contents for Synergy Compact
Page 14: ...Synergy Controller Technical Manual Revision H Page 14 2 0 SPECIFICATIONS 2 1 Data Sheet...
Page 71: ...Synergy Controller Technical Manual Revision H Page 71...
Page 106: ...Synergy Controller Technical Manual Revision H Page 106...
Page 123: ...Synergy Controller Technical Manual Revision H Page 123...
Page 141: ...Synergy Controller Technical Manual Revision H Page 141 TE1151 6 Triac Output Board Schematic...
Page 142: ...Synergy Controller Technical Manual Revision H Page 142 TE1708 6 Relay Board Schematic...
Page 188: ...Synergy Controller Technical Manual Revision H Page 188...
Page 274: ...Synergy Controller Technical Manual Revision H Page 274 Synergy Controller Step Syntax...
Page 294: ...Synergy Controller Technical Manual Revision H Page 294...
Page 312: ...Synergy Controller Technical Manual Revision H Page 312...
Page 345: ...Synergy Controller Technical Manual Revision H Page 345...
Page 346: ...Synergy Controller Technical Manual Revision H Page 346...