Synergy Controller Technical Manual, Revision H
Page 205
Vent – Boost Cool
This output specifies to use either Vent or Boost Cool.
Output is On/Off
Control Parameters
Default Value:
Range:
OT18
Selects the Control logic for the Vent – Boost Cool device
L14 Time Delay Boost Cool
Time delay (in seconds) required before Boost Cool is
enabled
Notes:
Turn On Timer is reset when Channel 1 PID Cool < 100%
PID Humidify
This output energizes the humidity generator and controls water vapor injection into the chamber.
Output is Time Proportioning
Control Parameters
Default Value:
Range:
Channel 2 Proportional Band Heating
42
0 - 50 Degrees C
Reset for Channel 2 Heating
0.02
0 - 9.9 Repeats/minute
Rate for Channel 2 Heating
0
0 - 9.9 Minutes
Cycle Time for Channel 2 Heating
1
1 - 60 Seconds
Rate Band for Channel 2 Heating
0
0 - 7 Seconds
Summary of Contents for Synergy Compact
Page 14: ...Synergy Controller Technical Manual Revision H Page 14 2 0 SPECIFICATIONS 2 1 Data Sheet...
Page 71: ...Synergy Controller Technical Manual Revision H Page 71...
Page 106: ...Synergy Controller Technical Manual Revision H Page 106...
Page 123: ...Synergy Controller Technical Manual Revision H Page 123...
Page 141: ...Synergy Controller Technical Manual Revision H Page 141 TE1151 6 Triac Output Board Schematic...
Page 142: ...Synergy Controller Technical Manual Revision H Page 142 TE1708 6 Relay Board Schematic...
Page 188: ...Synergy Controller Technical Manual Revision H Page 188...
Page 274: ...Synergy Controller Technical Manual Revision H Page 274 Synergy Controller Step Syntax...
Page 294: ...Synergy Controller Technical Manual Revision H Page 294...
Page 312: ...Synergy Controller Technical Manual Revision H Page 312...
Page 345: ...Synergy Controller Technical Manual Revision H Page 345...
Page 346: ...Synergy Controller Technical Manual Revision H Page 346...