Specifications
8
8.1
Acquisition system
Number of input channels
4 analog
CH1, CH2, CH3, CH4
BNC
Type
Single ended
Resolution
12, 14, 16 bit user selectable
Accuracy
0.2% of full scale
±
1 LSB
Ranges (full scale)
±
200 mV
±
2 V
±
20 V
±
400 mV
±
4 V
±
40 V
±
800 mV
±
8 V
±
80 V
Coupling
AC/DC
Impedance
1 M
Ω
/ 30 pF
Maximum voltage
200 V (DC + AC peak <10 kHz)
Bandwidth (-3dB)
50 MHz
AC coupling cut off frequency (-3dB)
±
1.5 Hz
Maximum sampling rate
HS4-50
HS4-25
HS4-10
HS4-5
12 bit
50 MSa/s
25 MSa/s
10 MSa/s
5 MSa/s
14 bit
3.125 MSa/s
3.125 MSa/s
3.125 MSa/s
3.125 MSa/s
16 bit
195 kSa/s
195 kSa/s
195 kSa/s
195 kSa/s
Maximum streaming rate
HS4-50
HS4-25
HS4-10
HS4-5
12 bit
500 kSa/s
250 kSa/s
100 kSa/s
50 kSa/s
14 bit
480 kSa/s
250 kSa/s
99 kSa/s
50 kSa/s
16 bit
195 kSa/s
195 kSa/s
97 kSa/s
48 kSa/s
Sampling source
internal quartz, external
Internal
Quartz
Accuracy
±
0.01%
Stability
±
100 ppm over -40
◦
C to +85
◦
C
External
On extension connector
Voltage
3.3 V TTL, 5 V TTL tolerant
Frequency range
95 MHz to 105 MHz
Memory
128 kSamples per channel
Specifications
23
Summary of Contents for Handyscope HS4
Page 1: ...Handyscope HS4 User manual TiePie engineering...
Page 6: ......
Page 17: ...Figure 4 3 Driver install Finished Driver installation 13...
Page 18: ...14 Chapter 4...
Page 20: ...16 Chapter 5...
Page 22: ...18 Chapter 6...
Page 26: ...22 Chapter 7...
Page 30: ...26 Chapter 8...
Page 32: ...TiePie engineering Handyscope HS4 instrument manual revision 2 27 February 2020...