9.3
Trigger system
System
Digital, 2 levels
Source
CH1, CH2, CH3, CH4, digital external, OR
Trigger modes
Rising edge, falling edge, any edge,
inside window, outside window,
enter window, exit window,
pulse width
Level adjustment
0 to 100% of full scale
Hysteresis adjustment
0 to 100% of full scale
Resolution
0.024 % (12 bits)/0.006 % (14/16 bits)
Pre trigger
0 to selected record length, 1 sample resolution
Post trigger
0 to selected record length, 1 sample resolution
Trigger hold-off
0 to 64 MSamples, 1 sample resolution
Trigger delay
0 to 16 GSamples, 1 sample resolution
Digital external trigger
Input
Extension connector pins 1 and 2
Range
0 to 2.5 V (TTL)
Coupling
DC
Jitter
≤
1 sample
9.4
Power
Power
From USB or external input
Consumption
5 V
DC
, 1200 mA max
Power adapter
External
Input
110 to 240 V
AC
, 50 to 60 Hz
0.85 A Max., 50 VA to 80 VA
Output
5.5 V
DC
, 2 A
Dimension
Height
30 mm / 1.2”
Width
45 mm / 1.8”
Length
75 mm / 3”
Order number
TP-UE15WCP1-055200SPA
Replaceable mains plugs for
EU, US, AU, UK
9.5
Multi-instrument synchronization
Maximum number of instruments
Limited by available number of USB ports
Synchronization accuracy
0 ppm
9.6
Probe calibration
Output
Extension connector pins 3 (signal) and 6 (ground)
Signal
Square wave
Level
-1 V to 1 V
Frequency
1 kHz
Specifications
35
Summary of Contents for Handyscope HS6 DIFF Series
Page 1: ...Handyscope HS6 DIFF User manual TiePie engineering...
Page 6: ......
Page 21: ...Figure 3 15 over compensated Introduction 17...
Page 22: ...18 Chapter 3...
Page 25: ...Figure 4 3 Driver install Finished Driver installation 21...
Page 26: ...22 Chapter 4...
Page 32: ...28 Chapter 7...
Page 44: ...TiePie engineering Handyscope HS6 DIFF instrument manual revision 2 27 February 2020...