background image

Description

1.

Transformer jaw
Used to pick up the current signal. To measure current, the jaw must be
firmly closed, leaving no air gap.

2.

Transformer trigger
Used to open the jaw.

3.

Range selector switch
Used to select the input range or switch the unit off.

4.

Zero button
This buton sets the current readout to zero. This function is also used when
measuring DC current to remove an offset value caused by residual mag-
netism in the core of the jaw.

5.

Power ON/OFF led.

6.

Low battery incidator
When the indicator lights, the battery should be replaced

7.

Output terminal – (black)
This terminal is used as output for DCA – polarity and ACA.

8.

Output te (red)
This terminal is used as output for DCA + polarity and ACA.

4

Current Clamp TP-CC80

Summary of Contents for TP-CC80

Page 1: ...Current Clamp TP CC80 User s manual Revision 1 3...

Page 2: ...conductors with voltages equal to or exceeding 300 V DC or 240 Vrms AC To avoid physical injury measurements on bare conductors or conductors with damaged insulation are not allowed Check the Current...

Page 3: ...80 for which this declaration is valid is in compliance with EN 55011 2009 A1 2010 EN 61000 6 1 2007 EN 55022 2006 A1 2007 EN 61000 6 3 2007 according the conditions of the EMC standard 2004 108 EC al...

Page 4: ...on This buton sets the current readout to zero This function is also used when measuring DC current to remove an o set value caused by residual mag netism in the core of the jaw 5 Power ON OFF led 6 L...

Page 5: ...om the upside to the underside of the clamp 6 Read the measured voltage value and convert it to the corresponding cur rent value 20 A range 100 mV equals 1 A 80 A range 100 mV equals 10 A AC current m...

Page 6: ...e test leads from the Current Clamp TP CC80 2 Remove the screw from the battery compartment at the back of the Current Clamp TP CC80 3 Remove the old battery 4 Insert a new 9 V NEDA 1604 IEC 6F22 batt...

Page 7: ...ronment 20 C to 70 C 80 RH Compliances CE Yes ROHS Yes Accuracy At 23 C 5 C 70 RH maximum DC 20 A range 3 5 mA DC 80 A range 100 mA to 40 A 3 0 3 A 40 A to 80 A 8 0 3 A AC 20 A range 10 mA to 10 A 40...

Page 8: ...7 8601 WL Sneek The Netherlands Tel 31 515 415 416 Fax 31 515 418 819 E mail sales tiepie nl www tiepie com This information is subject to change without notice Copyright 2018 TiePie engineering All r...

Reviews: