Operating instructions

DC current measurements

1.

Set the slide switch to 20 A DC or 80 A DC

2.

Connect the negative output of the Current Clamp TP-CC80 (black) to the
negative side of the input of the measuring instrument. Connect the positive
output of the Current Clamp TP-CC80 (red) to the positive side of the input
of the measuring instrument.

3.

Switch the measuring instrument on and set it to 200 mV DC, 400 mV DC or
2 V DC full scale input range.

4.

With firmly closed jaw, press the zero button until the instruments reads
zero.

5.

Open the jaw and fully enclose the current-carrying conductor(s) and close
the jaw again, no air gap is allowed between the two halves. The output
voltage is positive when the current flows from the upside to the underside
of the clamp.

6.

Read the measured voltage value and convert it to the corresponding cur-
rent value:
20 A range: 100 mV equals 1 A
80 A range: 100 mV equals 10 A.

AC current measurements

1.

Set the slide switch to 20 A AC or 80 A AC

2.

Connect the negative output of the Current Clamp TP-CC80 (black) to the
negative side of the input of the measuring instrument. Connect the positive
output of the Current Clamp TP-CC80 (red) to the positive side of the input
of the measuring instrument.

3.

Switch the measuring instrument on and set it to 200 mV AC, 400 mV AC or
2 V AC full scale input range.

4.

Open the jaw and fully enclose the current-carrying conductor(s) and close
the jaw again, no air gap is allowed between the two halves.

5.

Read the measured voltage value and convert it to the corresponding cur-
rent value:
20 A range: 100 mV equals 1 A
80 A range: 100 mV equals 10 A.

Current Clamp TP-CC80

5

Summary of Contents for TP-CC80

Page 1: ...Current Clamp TP CC80 User s manual Revision 1 3...

Page 2: ...conductors with voltages equal to or exceeding 300 V DC or 240 Vrms AC To avoid physical injury measurements on bare conductors or conductors with damaged insulation are not allowed Check the Current...

Page 3: ...80 for which this declaration is valid is in compliance with EN 55011 2009 A1 2010 EN 61000 6 1 2007 EN 55022 2006 A1 2007 EN 61000 6 3 2007 according the conditions of the EMC standard 2004 108 EC al...

Page 4: ...on This buton sets the current readout to zero This function is also used when measuring DC current to remove an o set value caused by residual mag netism in the core of the jaw 5 Power ON OFF led 6 L...

Page 5: ...om the upside to the underside of the clamp 6 Read the measured voltage value and convert it to the corresponding cur rent value 20 A range 100 mV equals 1 A 80 A range 100 mV equals 10 A AC current m...

Page 6: ...e test leads from the Current Clamp TP CC80 2 Remove the screw from the battery compartment at the back of the Current Clamp TP CC80 3 Remove the old battery 4 Insert a new 9 V NEDA 1604 IEC 6F22 batt...

Page 7: ...ronment 20 C to 70 C 80 RH Compliances CE Yes ROHS Yes Accuracy At 23 C 5 C 70 RH maximum DC 20 A range 3 5 mA DC 80 A range 100 mA to 40 A 3 0 3 A 40 A to 80 A 8 0 3 A AC 20 A range 10 mA to 10 A 40...

Page 8: ...7 8601 WL Sneek The Netherlands Tel 31 515 415 416 Fax 31 515 418 819 E mail sales tiepie nl www tiepie com This information is subject to change without notice Copyright 2018 TiePie engineering All r...

Reviews: