Safe
Ground protects your scope, your computer and your circuit under test against
accidental wrong ground connections. The
Safe
Ground protected ground feature
is optionally available for the WiFiScope WS6 DIFF.
4.1.2
Differential attenuators
To increase the input range of the WiFiScope WS6 DIFF, it comes with a differential
1:10 attenuator for each channel. This differential attenuator is specially designed
to be used with the WiFiScope WS6 DIFF.
Figure 4.5: Differential attenuator
For a differential input, both sides of the input need to be attenuated.
Figure 4.6: Differential attenuator matches with differential input
Standard oscilloscope probes and attenuators only attenuate one side of the sig-
nal path. These are not suitable to be used with a differential input. Using these
on a differential input will have a negative effect on the CMRR and will introduce
measurement errors.
Figure 4.7: Standard probe is mismatch with differential input
14
Chapter 4
Summary of Contents for WiFiScope WS6
Page 1: ...WiFiScope WS6 DIFF User manual USB Network WiFi Battery power TiePie engineering...
Page 6: ...IV...
Page 8: ...2 Chapter 1...
Page 10: ......
Page 14: ...8 Chapter 3...
Page 27: ...Figure 4 15 over compensated Introduction 21...
Page 28: ...22 Chapter 4...
Page 31: ...Figure 5 3 Driver install Finished Driver installation 25...
Page 32: ...26 Chapter 5...
Page 58: ...TiePie engineering WiFiScope WS6 DIFF instrument manual revision 2 27 February 2020...