Figure 8.2: CMI connectors
Connecting is done by daisy chaining the CMI connectors of the instruments prior
to starting the software, using special coupling cables (order number TP-C50H).
The software will detect how the instruments are connected to each other and
will automatically terminate the connection bus. The software will combine the
connected instruments to one large instrument. The combined instruments will
sample using the same clock, with a deviation of 0 ppm.
Figure 8.3: Multiple WiFiScope WS6 DIFFs combined
A 20 channel instrument is easily made by connecting five WiFiScope WS6 DIFFs
to each other.
When combining one or more WiFiScope WS6 DIFFs with other instruments includ-
ing Handyscope HS5s and/or WiFiScope WS5s, the daisy chained CMI bus must
begin or end with a Handyscope HS6 DIFF, WiFiScope WS6, WiFiScope WS6 DIFF,
Automotive Test Scope ATS610004D-XMSG, Automotive Test Scope ATS605004D-
XMS, Automotive Test Scope ATS610004DW-XMSG or Automotive Test Scope ATS605004DW-
XMS. Additionally, the maximum sampling rate is limited to 100 MSa/s at 14 bit
resolution.