Cannot be used in combination with a standard attenuating oscilloscope
probe
4.1.1
Safe
Ground on each channel
The
Safe
Ground protected ground feature of the WiFiScope WS6 DIFF allows to
switch the differential inputs to single ended inputs. This enables to perform single
ended measurements as well as using (attenuating) oscilloscope probes.
Safe
Ground
can be enabled for each channel individually.
Figure 4.3: SafeGround
A button on each channel toolbar in the software allows to control the
Safe
Ground
state between differential
and single ended
When
Safe
Ground is enabled for a channel, the input is switched to single ended
and the ground current of that input is real time monitored. When the ground
current becomes too high (<500 mA), the ground connection is immediately (<100
ns) opened, protecting the input against short circuits to ground. Also a warning
dialog is shown:
Figure 4.4: SafeGround warning
Once the short circuit is removed,
Safe
Ground can be enabled again.