5 Operation
SmartVap Vaporizer SW 1.n, Instructions for use (en-US)
49
A
Check that the locking knob (1) is unlocked.
► The locking knob is in horizontal position.
B
Check that each plug connector on the connector panel has an intact sealing ring, and
that there are no foreign bodies on the plug connectors.
C
Hold the vaporizer in an upright position with both hands. Place it carefully onto the con-
nector panel so that the plug connectors on the connector panel fit exactly into the
openings in the vaporizer.
D
Turn the locking knob (1) to the left (clockwise).
► The locking knob (1) is in vertical position. The vaporizer is locked (Figure b).
E
Ensure that the vaporizer is properly connected and locked by attempting to lift it off. If
the vaporizer can be lifted off, perform the connection once again.
1
1
b. Vaporizer locked
a. Vaporizer not locked