SmartVap Vaporizer SW 1.n, Instructions for use (en-US)
5.5
Filling
The vaporizer is filled through the Piramal Fill filling system, using Piramal DESFLURANE
bottles:
when disconnected from the mains,
in Standby mode (screen is turned off),
with the vaporizer turned on.
When the vaporizer is on, the fill level is displayed on the screen by the fill level indicator.
Please note:
The vaporizer may also be filled while dispensing Desflurane.
CAUTION
Risk to patient from underdosing of volatile anaesthetic, result-
ing from a reduced dispensing quantity due to filling being per-
formed during the delivery of anaesthesia.
With a high dispensing quantity (e.g. during uptake) do not fill
the reservoir.
CAUTION
Risk to persons from contamination of the environment with
volatile anaesthetic.
Avoid spilling liquid volatile anaesthetic.
Never press down manually on the filling valve under the
cover flap of the fill opening.
Ensure proper ventilation of the premises while filling.
IMPORTANT Damage to the device may be caused by the use of bottle
adapters
not compatible with the vaporizer’s filling system.
Only use Desflurane bottles with Piramal Fill.
Summary of Contents for SmartVap D
Page 1: ...Instructions for use SmartVapTM D SmartVapTM S SW 1 n...
Page 2: ......
Page 3: ...SmartVap Vaporizer SW 1 n Instructions for use en US 3 Page intentionally left blank...
Page 126: ...17 PIN information 126 SmartVap Vaporizer SW 1 n Instructions for use en US...
Page 128: ...128 SmartVap Vaporizer SW 1 n Instructions for use en US Notes...
Page 129: ...SmartVap Vaporizer SW 1 n Instructions for use en US 129 Notes...
Page 130: ...130 SmartVap Vaporizer SW 1 n m Instructions for use en US Notes...
Page 131: ......