5 Operation
58
SmartVap Vaporizer SW 1.n, Instructions for use (en-US)
D
Confirm the set dispensing quantity of Desflurane by briefly pressing the turning knob
(2).
► The illuminated ring around the turning knob shows a circling green point of light.
► The quantity of Desflurane dispensed into the fresh gas flow of the anaesthesia
workstation is changed.
If the dispensing quantity is set once again below the value of 12.5 Vol %, there will be no
limitation and no prompt for confirmation. The lower value can be set immediately and con-
firmed by pressing the turning knob.
i
Additional information about setting the dispensing quantity can be found in chapter 4.2.
5.8
Ending dispensing
A
Set the dispensing quantity to 0.0 by swiping right on the indicated screen area (1) or by
turning the knob (2) in a counterclockwise direction.
► The illuminated ring around the turning knob flashes blue.
B
Confirm the set value of 0.0 by briefly pressing the turning knob (2).
► The illuminated ring around the turning knob goes out.
► The dispensing of Desflurane into the fresh gas flow is ended.
i
Additional information about setting the dispensing quantity can be found in chapter 4.2.
1
2
Summary of Contents for SmartVap D
Page 1: ...Instructions for use SmartVapTM D SmartVapTM S SW 1 n...
Page 2: ......
Page 3: ...SmartVap Vaporizer SW 1 n Instructions for use en US 3 Page intentionally left blank...
Page 126: ...17 PIN information 126 SmartVap Vaporizer SW 1 n Instructions for use en US...
Page 128: ...128 SmartVap Vaporizer SW 1 n Instructions for use en US Notes...
Page 129: ...SmartVap Vaporizer SW 1 n Instructions for use en US 129 Notes...
Page 130: ...130 SmartVap Vaporizer SW 1 n m Instructions for use en US Notes...
Page 131: ......