5 Operation
SmartVap Vaporizer SW 1.n, Instructions for use (en-US)
59
5.9
Standby mode
If the vaporizer is not being used, it can be put into Standby mode. In Standby mode, the
vaporizer continues to be heated and the emergency battery continues to be charged.
5.9.1
Opening Standby dialog
Prerequisite:
The vaporizer is in the operating screen and dispensing of Desflurane
has ended.
There are two ways to open the
Standby
dialog:
Option 1: Via the touchscreen and Menu mode
A
Tap on the
Menu mode
(1) button in the screen’s bottom bar.
► The Menu bar opens and the
Case data
dialog is displayed.
B
Tap the
Standby
(2) button in the Menu bar.
► The
Standby
dialog is displayed.
i
Additional information about the Menu mode can be found in chapter 7.
1
2