background image

 

 

 
 
 
 

User Manual 

 

5068 InsCal 

Insulation Tester Calibrator

 

 

Version 2.5

 

11-22

 

 
 
 
 

 

 

 
 
 

Time Electronics Ltd

 

 

Unit 5, TON Business Park, 2-8 Morley Road,  
Tonbridge, Kent, TN9 1RA, United Kingdom.   
T: +44 (0) 1732 355993  |  F: +44 (0) 1732 350198 
mail@timeelectronics.co.uk  |  www.timeelectronics.com 

Summary of Contents for 5068 Ins-Cal

Page 1: ...tion Tester Calibrator Version 2 5 11 22 Time Electronics Ltd Unit 5 TON Business Park 2 8 Morley Road Tonbridge Kent TN9 1RA United Kingdom T 44 0 1732 355993 F 44 0 1732 350198 mail timeelectronics...

Page 2: ...Electronics Ltd This also applies to any schematics drawings and diagrams contained herein This manual provides operating and safety instructions for the Time Electronics product To ensure correct op...

Page 3: ...pecification 6 3 3 Ordering Information 6 4 Controls 7 4 1 Descriptions of Controls 7 5 Operation 8 5 1 Operational Guidelines 8 5 2 Typical Operation Procedure 9 5 2 1 Initial Setup 9 5 2 2 Voltage M...

Page 4: ...rom any external circuits including the mains supply The switch in the insulation resistance selection circuit is constructed from special high voltage low leakage wafers and are self cleaning so shou...

Page 5: ...cs 5068 Insulation Tester Calibrator User Manual 5 2 Supplied Items 5068 InsCal Insulation Tester Calibration System Accessories supplied Test Lead Wallet Low resistance test leads 2 5 kV test leads B...

Page 6: ...ull scale reading High Range 0 to 20 mA 1 of full scale reading Terminal impedance 447 this includes the discharge resistor for short circuit test 3 2 General Specification Power Internal battery 6 V...

Page 7: ...he V or I measure function Function selector and insulation resistance setting knob Use to select the required function or the insulation resistance spot values from 100 k to 10 G Low resistance conne...

Page 8: ...interference which can cause a disturbance to the readings obtained Some manufactures fit a guard connection for use with dedicated leads this will stabilise the readings It is not recommended to oper...

Page 9: ...1 Check the battery on both 5068 V I Meter and Insulation Tester Make sure their batteries are fully charged 2 Complete the setup procedure for the insulation tester UUT 3 Connect the selected test le...

Page 10: ...he UUT Analogue meter display if required 2 On the 5068 select the O C VOLTS function using the rotator knob 3 Using range selection button under the display choose the required voltage range of 1 999...

Page 11: ...he S C CURRENT function using the rotator knob 2 Using range selection button under the display choose the required current range 2 mA or 20 mA 3 Press the test button on the UUT and record the curren...

Page 12: ...ot value resistances shown on the 5068 panel Ensure that the voltage set on the UUT does not exceed this voltage rating Important Note Wherever possible use the test leads supplied with the UUT 5 Pres...

Page 13: ...of the 5068 in conjunction with the supplied low voltage test lead set the continuity feature of the UUT can be tested and verified The low resistance accuracy of the UUT can also be verified Connect...

Page 14: ...e user such as misuse operation outside of specification improper maintenance or repair or unauthorized modification Time Electronics total liability is limited to repair or replacement of the product...

Page 15: ...e Electronics Ltd Unit 5 TON Business Park 2 8 Morley Road Tonbridge Kent TN9 1RA United Kingdom Tel 44 0 1732 355993 Fax 44 0 1732 350198 Email mail timeelectronics co uk Web Site www timeelectronics...