Tinytag Plus Radio Data Logger with Temperature & Relative Humidity Probe
Tinytag Plus Radio Data Logger with Temperature & Relative Humidity Probe
Tinytag Plus Radio Data Logger with Temperature & Relative Humidity Probe
Tinytag Plus Radio Data Logger with Temperature & Relative Humidity Probe
((((----25 to +85°C/0 to 100% RH)
25 to +85°C/0 to 100% RH)
25 to +85°C/0 to 100% RH)
25 to +85°C/0 to 100% RH)
TGRF-4600
Issue 1
Issue 1
Issue 1
Issue 1 (19th April 2019) E&OE
www.tinytag.info
www.tinytag.info
www.tinytag.info
www.tinytag.info
sales@tinytag.info
sales@tinytag.info
sales@tinytag.info
sales@tinytag.info
Reading Specification
Connections
Physical Specification
IP Rating
IP67
Operational Range*
-20°C to +55°C
Case Dimensions (excluding antenna)
Length/Height
241mm / 9.49"
Width
155mm / 6.10"
Depth
62mm / 2.44"
Weight
(inc. antenna)
554g / 19.54oz
*The Operational Range indicates the physical limits to which the unit can be
exposed.
Temperature
Reading Range
-25°C to +85°C
Sensor Type
10K NTC Thermistor
Response Time
5mins to 90% of the reading change
(in moving air)
Reading Resolution
0.01°C or better
Accuracy
Relative Humidity
Reading Range
0 to 100% RH
Accuracy
±3.0% RH
Reading Resolution
0.04% RH
Response Time
90 seconds to 90% of the reading change
(in moving air)
Stability
Better than 1% per year (typical)
If moisture forms on the RH sensor readings may become unpredictable.
Any dust or residue that is allowed to build up on the RH sensor will affect the
reading accuracy.
The sensor may be cleaned with compressed air.
A
PBRF-8002
temperature and relative humidity probe and
CAB-0034
sensor
extension cable are provided with the data logger.
IP Rating
IP20
Operational Range*
-25°C to +85°C
Dimensions
Probe Length (excluding cable)
61mm / 2.4”
Maximum Diameter
16mm / 0.63”
Cable Length
1.5m / 59.055’
Weight
10g / 0.35oz
*The Operational Range indicates the physical limits to which the probe and
cable can be exposed.