Potential equalization
Requirements
Please consider the following to ensure correct measurement:
• Same electrical potential for the fluid and sensor
• Remote version: same electrical potential for the sensor and transmitter
• Company-internal grounding concepts
• Pipe material and grounding
Terminals
• For device version without integrated overvoltage protection: plug-in spring terminals for wire
cross-sections 0.5 to 2.5 mm
2
(20 to 14 AWG)
• For device version with integrated overvoltage protection: screw terminals for wire cross-sections
0.2 to 2.5 mm
2
(24 to 14 AWG)
Cable entries
• Thread for cable entry:
• For non-hazardous and hazardous areas (not for XP): G ½"
Cable specification
Permitted temperature range
• The installation guidelines that apply in the country of installation must be observed.
• The cables must be suitable for the minimum and maximum temperatures to be expected.
Signal cable
Current output 4 to 20 mA
A shielded cable is recommended. Observe grounding concept of the plant.
Pulse/frequency/switch output
Standard installation cable is sufficient.
Connecting cable for remote version
Connecting cable (standard)
Standard cable
2 × 2 × 0.5 mm
2
(22 AWG) PVC cable with common shield (2 pairs, pair-
stranded)
1)
Flame resistance
According to DIN EN 60332-1-2
Oil-resistance
According to DIN EN 60811-2-1
Shielding
Galvanized copper-braid, opt. density approx. 85%
Cable length
5 m (16 ft), 10 m (32 ft), 20 m (65 ft), 30 m (98 ft)
Operating temperature
When mounted in a fixed position: ‒50 to +105
°
C (‒58 to +221
°
F);
when cable can move freely: ‒25 to +105
°
C (‒13 to +221
°
F)
1)
UV radiation may cause damage to the outer jacket of the cable. Protect the cable from exposure to sun
as much as possible.