EF200R-C
39
Nominal pressure of sensor
The following overpressure resistance values apply to the sensor shaft in the event of a membrane
rupture:
Sensor version; DSC sensor; measuring tube
Overpressure, sensor shaft
in [bar a]
Mass (integrated temperature measurement)
200
Mass steam (integrated pressure/temperature measurement)
200
Pressure specifications
The OPL (over pressure limit = sensor overload limit) for the measuring device depends on the
lowest-rated element, with regard to pressure, of the selected components, i.e. the process
connection has to be taken into consideration in addition to the measuring cell. Also observe
pressure-temperature dependency. For the appropriate standards and further information →
See page 26. The OPL may only be applied for a limited period of time.
The MWP (maximum working pressure) for the sensors depends on the lowest-rated element, with
regard to pressure, of the selected components, i.e. the process connection has to be taken into
consideration in addition to the measuring cell. Also observe pressure-temperature dependency. For
the appropriate standards and further information → See page 26. The MWP may be applied
at the devicefor an unlimited period. The MWP can also be found on the nameplate.
Flange connection: flange according to JIS B2220
0
10
20
30
40
50
60
-200
-100
0
100
200
300
400
[°C]
[°F]
[bar]
[psi]
0
100
200
300
400
500
600
700
800
900
-400
-200
0
800
200
600
400
20K
10K
0
10
20
30
40
50
60
-200
-100
0
100
200
300
400
[°C]
[°F]
[bar]
[psi]
0
100
200
300
400
500
600
700
800
900
-400
-200
0
800
200
600
400
20K
10K
A0034043-EN
Flange connection material: stainless steel, multiple certifications, 1.4404/F316/F316L