Model: TTT-710
4.3 Piping
- 27 -
⑥
Insert the plug.
⋅⋅⋅⋅⋅⋅
Insert the plug of liquid level sensor into the liquid level sensor connection
connector (for pure water) (LEVEL ALARM RINSE 1) of the product main body.
Inserting the Plug of Liquid Level Sensor (for pure water)
(3) Waste solution tank connection
①
Cut the waste solution hose.
⋅⋅⋅⋅⋅⋅
Use a hose cuter, etc. to cut the accessory waste solution
hose (part code: XD701500) to a suitable length matching to the distance between waste solution
outlet 2 of product main body and the waste solution tank to be used and use the hose as the waste
solution hose (for outlet 2).
②
Connect the waste solution hose.
⋅⋅⋅⋅⋅⋅
Connect one end of the waste solution hose (for outlet
2) prepared at step
①
to waste solution outlet 2 of product main body and the other end to the
waste solution tank piping port.
【
IMPORTANT
】
•
Install the waste solution tank at a place sufficiently lower than the
product main body. Also be sure that the waste solution hose (for outlet 2)
is straight from top to bottom. If the hose is installed at a high place or is
bent midway between top and bottom, the waste solution will collect in the
hose and will not flow into the waste solution tank.