Model: TTT-710
6.1 Manual Operation
- 64 -
(Continued from previous page)
Operation Screen
example
②
Perform the washing operation.
⋅⋅⋅
Press
or
and select “1. Start Washing” at the Electrode Washing
Selection screen and press
ENTER
.
•
The Washing Operating screen appears.
•
The electrode arm moves to the washing reservoir and
washing is performed in accordance with the conditions set
at 6.2(7) “Electrode washing conditons setting.
•
During operation, the executing operation sequence is
displayed.
Washing Operating screen
(Pure water shower washing)
③
End the washing operation.
⋅⋅⋅
When the washing
operation ends, after the electrode arm stops above the
washing reservoir, the End of Washing screen is displayed
automatically.
•
The electrode arm movement, table rotation, stirring, and
re-washing operations can be performed at this screen.
: Arm up (When this key is pressed again while the arm
is at the topmost position, the arm moves to the left or
right.)
〔
Note
〕
•
When the arm is washing reservoir position, it
moves to the right. When the arm is in the table
position, it moves to the left.
: Arm down
: Table plate rotates 1 specimen.
STIRRER
: Stirring on/off
ENTER
: Washing operation re-execution
End of Washing screen
④
End the manual washing.
⋅⋅⋅
Press
WASH
at the step
③
screen.
•
The arm moves to the electrode storage beaker and the
table moves to the home position and operation ends.
•
When
WASH
is pressed during the stirring operation, after
stirring stops, operation can be ended the same as
mentioned above.
•
In any case, the display returns to the Measurement
Waiting screen.