Model: TTT-710
7.2 Message Format
- 84 -
When request command of tank liquid level sensor status was received
QS04,x x xCRLF
① ②③④
①
Identification code QS04
②
Washing water 1 (pure water)
0: Normal 1: Abnormal (washing water empty)
③
Washing water 2 (reagent)
0: Normal 1: Abnormal (washing water empty)
④
Waste solution
0: Normal 1: Abnormal (waste solution full)
When request command of current table No. was received
QS05,xxxCRLF
① ②
①
Identification code QS05
②
Current table No.: 001 to 100
When request command of end pin detection status was received
QS06,xCRLF
① ②
①
Identification code QS06
②
End pin detection status 0: End pin not detected 1: End pin detected
When request command of washing mode No. was received
Washing mode 1
QW1,xxxx,xCRLF
① ② ③
①
Identification code QW1
②
Pure water shower time: 0000 to 9999
③
Number of air blows: 1 to9
Washing mode 2
QW2,xxxx,xxxx,xCRLF
① ② ③ ④
①
Identification code QW2
②
Pure water bubbling time: 0000 to 9999
③
Pure water shower time: 0000 to 9999
④
Number of air blows: 1 to9