9.3 Inclinometer adapter
9-9
Note
1. From the monitor screen, go to “Channel
setting” and then to “Measurement mode”.
■
Measurement mode setting
Select Inclinometer 1-axis mode or Inclinometer 2-axis mode. For more detail of setting
procedures, refer to the page 8 of Chapter 4, “Switching of measurement mode”.
■
Channel setting
In 2-axis mode, an even number is set as first axis, and a sequential number is set as second axis,
like “00-01” or “02-03”. In 1-axis mode, any number from 00 to 19 can be set.
■ Monitor display
In 2-axis mode, two channels are simultaneously monitored and displayed.
1. In case of 2-axis mode, input any even
number from 00 to 18 in the cursor point,
and press
[ENT]
key.
In case of 1-axis mode, input any number
from 00 to 19 in the cursor point, and press
[ENT]
key.
2. Select “Inclinometer 1 axis mode” or
“Inclinometer 2 axes mode”.
Waveform display is not available
in inclinometer mode.
Summary of Contents for TC-32K
Page 1: ...O p e r a t i o n M a n u a l TC 32K HANDHELD DATA LOGGER...
Page 8: ...Chapter 12 Error Message 12 1 Explanations and countermeasures for error messages 12 2...
Page 9: ...Chapter 1 Overview 1 1 Overview 1 2 1 2 Features 1 2 1 3 Details about each part 1 3...
Page 22: ...2 5 Operation outline 2 10 memo...
Page 23: ...Chapter 3 Sensor Connection 3 1 Sensor connection 3 2...
Page 65: ...5 10 Measurement auxiliary setting 5 28 memo...
Page 78: ...6 5 Recording in data memory and CF card 6 13 memo...
Page 86: ...7 7 Remote measurement 7 8 memo...
Page 98: ...8 7 Factory setting 8 12 memo...
Page 127: ...11 4 Outside drawing 11 8 11 4 Outside drawing Unit mm...
Page 128: ...Chapter 12 Error Message 12 1 Explanations and countermeasures for error messages 12 2...