4.2 Monitor display
4-3
○
Example of 5-channel display (multi-channel mode)
■ Waveform monitor
The screen displays a line chart showing variation of measurement value with time. The number
of channel that can be displayed in the waveform monitor is one.
■
Meaning of displayed value
When the sensor is connected to the input terminal a nd the sensor and the measurement system
have no problem, a certain value is always displayed. However, if the sensor is not connected,
or connected but its wire is broken, or the input value is over the measurement range defined in
the specifications, the following indications are displayed to alert, instead of measurement values.
The indication that appears when a problem occurs, and supposed cause are shown below:
[ * * * * * * * ]
Open data
The sensor is not connected, or the sensor cable is broken.
[ + * * * * * * * ]
+Over data
The input value is over the measurement range in the + direction.
[ - * * * * * * * ]
-Over data
The input value is over the measurement range in the - direction.
[ * * * * I * * * ]
Initial value over data
The initial value (initial unbalanced value) is over the initial value storing range.
[ ! ! ! ! ! ! ! ]
Indication digits over
The indication digits (before the decimal point) are insufficient for displaying the
measurement value.
[
――JUMP――
]
Jump
This message is displayed when the sensor mode is “JUMP”.
Box number
CSW-5B channel
Mark showing that
the channel is
monitored
Measurement value
(monitored value)
Monitored channel
Summary of Contents for TC-32K
Page 1: ...O p e r a t i o n M a n u a l TC 32K HANDHELD DATA LOGGER...
Page 8: ...Chapter 12 Error Message 12 1 Explanations and countermeasures for error messages 12 2...
Page 9: ...Chapter 1 Overview 1 1 Overview 1 2 1 2 Features 1 2 1 3 Details about each part 1 3...
Page 22: ...2 5 Operation outline 2 10 memo...
Page 23: ...Chapter 3 Sensor Connection 3 1 Sensor connection 3 2...
Page 65: ...5 10 Measurement auxiliary setting 5 28 memo...
Page 78: ...6 5 Recording in data memory and CF card 6 13 memo...
Page 86: ...7 7 Remote measurement 7 8 memo...
Page 98: ...8 7 Factory setting 8 12 memo...
Page 127: ...11 4 Outside drawing 11 8 11 4 Outside drawing Unit mm...
Page 128: ...Chapter 12 Error Message 12 1 Explanations and countermeasures for error messages 12 2...