Safety Precautions
Powder or dust inside the system may cause poor contact or a lowered
insulation effect in the connector. Pay special attention, during use and
storage, not to allow dust to enter the system.
Do not operate the system in a place where there is flammable gas or
flammable steam. This may cause fire.
Danger
Operate the system at a specified temperature. If the operating site is
exposed to direct sunlight or an extremely low temperature, arrange for shade
or a thermal insulating material.
It is not recommended for the user to disassemble or remodel the system.
Such a do-it-yourself action may cause an electric shock or a malfunction.
Never connect the grounding cable to a gas pipe. In addition, make it a rule
to disconnect the power supply cable before connecting or disconnecting the
grounding cable. There is danger of a fire and the electric shock.
Operate the system at a relative humidity less than 85%. Do not expose it to
rain or extreme humidity. When water flowed into the system or system is
flooded, dry enough before turning on the power supply. If the system does
not start normally, some trouble might be caused. Please contact us.
Warning
The system may malfunction if either the unit or its wiring is placed near such
machines as a large motor, crane, transformer, or welding machine. When
extending the sensor to a place subject to a strong electric field, such as near
a power substation or radio transmission station, use a special cable such as
a shielded cable.
The system is vulnerable to the dielectric effect of thunderbolts. Take
preventive measures against thunderbolts where applicable. Contact your
dealer or Tokyo Sokki Kenkyujo for details.
Danger
Danger
Caution
Caution
Caution
Caution
Do not forcibly drag the system with connection cables connected. The
cables may be damaged and/or the connectors may be disconnected. In
addition, do not apply impact to the connector of the cable. Keep the
connectors away from soil, mud, water and oil.
Caution
The range of power supply voltage for this system is DC 9
~
18 V.
Malfunction may be caused if larger voltage is supplied. If the system is
used in a positive grounded vehicle, do not contact the housing of this system
with the body of the vehicle.
Caution
Summary of Contents for TC-32K
Page 1: ...O p e r a t i o n M a n u a l TC 32K HANDHELD DATA LOGGER...
Page 8: ...Chapter 12 Error Message 12 1 Explanations and countermeasures for error messages 12 2...
Page 9: ...Chapter 1 Overview 1 1 Overview 1 2 1 2 Features 1 2 1 3 Details about each part 1 3...
Page 22: ...2 5 Operation outline 2 10 memo...
Page 23: ...Chapter 3 Sensor Connection 3 1 Sensor connection 3 2...
Page 65: ...5 10 Measurement auxiliary setting 5 28 memo...
Page 78: ...6 5 Recording in data memory and CF card 6 13 memo...
Page 86: ...7 7 Remote measurement 7 8 memo...
Page 98: ...8 7 Factory setting 8 12 memo...
Page 127: ...11 4 Outside drawing 11 8 11 4 Outside drawing Unit mm...
Page 128: ...Chapter 12 Error Message 12 1 Explanations and countermeasures for error messages 12 2...