107
6.4.2.16
Address 17 (Frequency Low Limit)
1)
Commands Description
This command is used to set the low limit of the output frequency in manual mode. When the set voltage
is not greater than 150V, the command data range is 45.0~500Hz. When the frequency is less than 100 Hz,
the resolution is 0.1 Hz; when the frequency is not less than 100 Hz, the resolution is 1 Hz.
2)
Send Instruction Format
Instrument
address
Function
code
High
address
Low
address
High
bytes
Low
bytes
Number
of data
Data byte 1~4
CRC
Low
CRC
High
1~32
0x10
0x00
0x11
0x00
0x02
0x04
45.5~500, 4 bytes
stored in float
point type
Calculated based on
the first 11 bytes
3)
Return Instruction Format
6.4.2.17
Address 18 (Start Angle)
1)
Commands Description
This command is used to set the low limit of the output frequency in manual mode. When the set
This command is used to set the starting angle in manual mode. The command data range is 0~359. The
command is invalid when the output is turned on.
2)
Send Instruction Format
Instrument
address
Function
code
High
address
Low
address
High
bytes
Low
bytes
Number
of data
Data
byte 1
Data
byte 2
CRC
Low
CRC
High
1~32
0x10
0x00
0x12
0x00
0x01
0x02
0~359
High
8-bit
0~359
low
8-bit
Calculated based
on the first 9 bytes
3)
Return Instruction Format
Instrument
address
Function
Code
High Address
Low Address
High bytes
Low bytes
CRC Low
CRC High
1~32
0x10
0x00
0x11
0x00
0x02
Calculated based on the first 6
bytes
Instrument
address
Function
Code
High Address
Low Address
High bytes
Low bytes
CRC Low
CRC High
1~32
0x10
0x00
0x12
0x00
0x01
Calculated based on the first 6
bytes