112
1~32
0x10
0x00
0x1A
0x00
0x01
0x01
0x00
0x00~0x63
Calculated based
on the first 9
bytes
3)
Return Instruction Format
6.4.2.26
Address 27 (Memory Sequence)
1)
Commands Description
This command is used to set the number of the memory sequence in the programmable mode.
Instruction data byte 1 ranges from 1 to 50. The command is invalid when the output is turned on.
2)
Send Instruction Format
Instrument
address
Function
code
High
address
Low
address
High
bytes
Low
bytes
Number
of data
Data
byte 1
Data byte
2
CRC
Low
CRC
High
1~32
0x10
0x00
0x1B
0x00
0x01
0x01
0x00
0x01~0x32
Calculated based
on the first 8
bytes
3)
Return Instruction Format
6.4.2.27
Address 28 (Sequence Number)
1)
Command Description
This command is used to set the number of cycles of the current memory sequence in the programmed
mode. Instruction data range is 0~999.Command is invalid when output is turned on.
2)
Send Instruction Format
Instrument
address
Function
code
High
address
Low
address
High
bytes
Low
bytes
Number
of data
Data
byte 1
Data
byte 2
CRC
Low
CRC
High
Instrument
address
Function
Code
High Address
Low Address
High bytes
Low bytes
CRC Low
CRC High
1~32
0x10
0x00
0x1A
0x00
0x01
Calculated based on the first 6
bytes
Instrument
address
Function
Code
High Address
Low Address
High bytes
Low bytes
CRC Low
CRC High
1~32
0x10
0x00
0x1B
0x00
0x01
Calculated based on the first 6
bytes