Return Instruction Format
6.4.2.32
Address 33 (Current High Limit)
1)
Command Description
This command is used to set the current high limit of the current step in the programmed mode. When
the set voltage is not greater than 150V, the command data range is 0.000~8.400A; when the set voltage
is greater than 150V, the command data range is 0.000~4.200A. Command is invalid when output is
turned on
.
2)
Send Instruction Format
Instrument
address
Function
code
High
address
Low
address
High
bytes
Low
bytes
Number
of data
Data byte 1~4
CRC
Low
CRC
High
1~32
0x10
0x00
0x21
0x00
0x02
0x04
0.000~8.400 or
0.000~4.200, 4
bytes stored in
float point type
Calculated based on
the first 11 bytes
3)
Return Instruction Format
6.4.2.33
Address 34 (Command Low Limit)
1)
Command Description
This command is used to set the current low limit value of the current step in the programmed mode.
When the set voltage is not greater than 150V, the command data range is 0.000~8.400A; when the set
voltage is greater than 150V, the command data range is 0.000~4.200A. Command is invalid when output
Instrument
address
Function
Code
High Address
Low Address
High bytes
Low bytes
CRC Low
CRC High
1~32
0x10
0x00
0x20
0x00
0x01
Calculated based on the first 6
bytes
Instrument
address
Function
Code
High Address
Low Address
High bytes
Low bytes
CRC Low
CRC High
1~32
0x10
0x00
0x21
0x00
0x02
Calculated based on the first 6
bytes