120
6.4.2.41
Address 42 (PF Low Limit)
1)
Command Description
This command is used to set the power factor low limit value of the current step in the programmed mode.
Instruction data range is 0.000~1.000. Command is invalid when output is turned on
.
2)
Send Instruction Format
Instrument
address
Function
code
High
address
Low
address
High
bytes
Low
bytes
Number
of data
Data byte 1~4
CRC
Low
CRC
High
1~32
0x10
0x00
0x2A
0x00
0x02
0x04
0.0~1000, 4 bytes
stored in float
point type
Calculated based on
the first 11 bytes
3)
Return Instruction Format
6.4.2.42
Address 43 (Time Unit)
1)
Command Description
This command is used to set the unit of delay determination time and test time in the programmed mode.
Command is invalid when output is turned on. The instruction data byte bit corresponds to the function of
the instrument as shown below
:
Data Byte Result Display Mode
0x00
This command is used to set the delay judgment time and test time in
the programmed mode as seconds.
0x01
This command is used to set the delay judgment time and test time in
the programmed mode as minutes.
0x02
This command is used to set the delay judgment time and test time in
the programmed mode as hours.
2)
Send Instruction Format
Instrument
address
Function
code
High
address
Low
address
High
bytes
Low
bytes
Number
of data
Data
byte 1
Data byte
2
CRC
Low
CRC
High
1~32
0x10
0x00
0x29
0x00
0x02
Calculated based on the first 6
bytes
Instrument
address
Function
Code
High Address
Low Address
High bytes
Low bytes
CRC Low
CRC High
1~32
0x10
0x00
0x2A
0x00
0x02
Calculated based on the first 6
bytes