129
This command is used to control the surge drop in the programmed mode. Command is invalid when
output is turned on. The command data byte bit corresponds to the function of the instrument as shown
below:
Data Byte
Step Connection
0x00
This command is used to turn off the surge drop in the program mode.
0x01
This command is used to turn on the surge drop in the program mode.
2)
Send Instruction Format
Instrument
address
Function
code
High
address
Low
address
High
bytes
Low
bytes
Number
of data
Data
byte 1
Data
byte 2
CRC
Low
CRC
High
1~32
0x10
0x00
0x3B
0x00
0x01
0x01
0x00
0x00
or
0x01
Calculated based
on the first 9 bytes
3)
Return Instruction Format
6.4.2.59
Address 60 (Over Current Fold)
1)
Command Description
This command is used to control the over current fold function in the programmed mode. Command is
invalid when output is turned on. The command data byte bit corresponds to the function of the
instrument as shown below:
Data Byte
over current fold function
0x00
This command is used to turn off the over current fold function in the programmed
mode.
0x01
This command is used to turn on the over current fold function in the programmed
mode.
2)
Send Instruction Format
Instrument
address
Function
code
High
address
Low
address
High
bytes
Low
bytes
Number
of data
Data
byte 1
Data
byte 2
CRC
Low
CRC
High
1~32
0x10
0x00
0x3E
0x00
0x01
0x02
0x00
0x00
or
0x01
Calculated based
on the first 8 bytes
Instrument
address
Function
Code
High Address
Low Address
High bytes
Low bytes
CRC Low
CRC High
1~32
0x10
0x00
0x3B
0x00
0x01
Calculated based on the first 6
bytes