162
6.4.3.61
Address 61 (Total Circle Numbers)
1)
Command Description
This command is used to query the total number of cycles of the memory sequence in the programmed
mode.
2)
Send Instruction Format
3)
Return Instruction Format
Return instruction data range is 0~999
6.4.3.62
Address 62 (Singe Step Test)
1)
Command Description
This command is used to query the single-step test status in the programmed mode.
2)
Send Instruction Format
3)
Return Instruction Format
Data byte returns 0 or 1. 0 means that the single-step test status is off in the programmed mode; 1 means
that the single-step test status is on in the programmed mode.
Instrument
address
Function
Code
High Address
Low Address
High bytes
Low bytes
CRC Low
CRC High
1~32
0x03
0x00
0x3D
0x00
0x01
Calculated based on the first 6
bytes
Instrument
address
Function
Code
Number of
Data
Data Byte1
Data Byte2
CRC Low
CRC High
1~32
0x03
0x02
0~999 High
8-bit
0~999 Low 8-bit
Calculated based on the first 5
bytes
Instrument
address
Function
Code
High Address
Low Address
High bytes
Low bytes
CRC Low
CRC High
1~32
0x03
0x00
0x3E
0x00
0x01
Calculated based on the first 6
bytes
Instrument
address
Function
Code
Number of
Data
Data Byte1
Data Byte2
CRC Low
CRC High
1~32
0x03
0x02
0x00
0x00
Calculated based on the first 5
bytes
0x01