parameters under the COMMON setting in <Setup> page. These parameter items are common to all
memory sequences under the MEMORY setting. As shown in Figure 3-9:
Figure3-9 General Setting<Manual Mode>
2) Operation
Press SETUP key, select <Manual Mode>, then make use of the knob and the
button to move the
cursor to the COMMON option.
3.2.2.2
Parameters Setting Description and Operation
V HiLmt
1)
Description
This parameter is used to set the high limit of output voltage. Its range is 0.0~300.0V.
2)
Operation
Make use of the
key or the knob to move the cursor to this parameter option, then change it
through the front panel numeric keys, then press ENTER or OK to confirm the change. If you do not
want to change the parameter value or press the wrong numeric key, you can cancel it by pressing the
key or ESC key.
V LoLmt
1)
Description
This parameter is used to set the low limit of output voltage. Its range is 0.0~300.0V.
2)
Operation
Make use of the
key or the knob to move the cursor to this parameter option, then change it
through the front panel numeric keys, then press ENTER or OK to confirm the change. If you do not
want to change the parameter value or press the wrong numeric key, you can cancel it by pressing the
key or ESC key.