26
BASIC OPERATIONS
BA
SI
C OPERATIONS
PREPARATION BEFORE M
EASUREM
E
NT
AIR CHECK
This instrument is equipped with a function for checking correct operations of the measurement sys-
tem inside the instrument.
1
Tap the
button.
Then, the confirming message of Air check/End operation is displayed.
2
Press the
button. Then, air is blown out from the measuring nozzle automatically.
3
Make sure the message box of "Air check. Press OK to continue." is displayed on the control
panel.
4
Press the
button, return to Measurement screen.
NOTE
If abnormal ) or abnormal action(-) is displayed, the condition is not
normal. Check if there is any object before the measuring nozzle.
If there is any, remove it, press the
button and do the check again.
If there is no object, a failure is suspected. Turn the
switch to
OFF, unplug the power cable, and call your dealer or TOPCON at the
address printed on the back cover of this manual.
Air check
Patient ID
Patient ID
1
1
TOPCON̲TAROU
TOPCON̲TAROU
Auto Shoot
Perform air check?
CANCEL
Check
CHECK
Air check. Press OK to continue.
OK
OK
POWER
Confirm abnormal action(-) of air check.