102
6. Input the information 3 and
press [ENT].
7. Collimate the staff in point a
and press [MEAS] (As shown in
the figure 1), meanwhile, take
measurement and the screen
will display Aa.
8. Collimate the staff in point b
and press [MEAS] (As shown in
the figure 1), meanwhile, take
measurement and the screen
will display Ab.
9. Move the instrument to
point B and level it (As shown
in the figure 2).
At this time, user can switch off
the instrument to save power.
10. Collimate the staff in point
a and press [MEAS] (As shown
in the figure 2), meanwhile,
take measurement and the
screen will display Ba.
Info 3
[ENT]
[
MEAS
]
[MEAS]
Adjust
Method A
Move A-------
→
B
Relocate
Adjust
Info #3
=>3
Adjust A
Point:A Rod:a
a
←
----A b
Press [MEAS] to meas
Adjust A
Point:A Rod:a
Aa Rod:1.0567m
N:3
δ
:0.02mm
Adjust A
Point:A Rod:b
a A----
→
b
Press [MEAS] to meas
Adjust A
Point:A Rod:b
Ab Rod:1.0567m
N:3
δ
:0.02mm