32
7. INSTRUMENT SETTINGS
By changing the scanning settings, measurement that suits the purpose can be carried out.
Resolution can be selected from the predetermined resolutions or by entering the desired resolution.
1. Press the Scan Setting icon.
• The Scan Setting icon displayed in the status screen
shows the selective condition of the range mode.
2. Select “Scan Area,” “Resolution,” “Mode,”
“Pulse Select,” and press [OK.]
(1) Scanning range
Range setting scanning or Whole circumference scanning can be selected.
“Scanning range” on page 64
(2) Resolution (a predetermined value)
Select a predetermined resolution listed below:
50mm@10m, 25mm@10m, 12.5mm@10m, 6.3mm@10m, 3.1mm@10m, Edit
“50mm@10 m” means “a measurement accuracy of 50mm in pitch at a position of 10 m ahead”.
For the range of measurement distance and the accuracy of measurement, see the following table.
*1) Range varies depending on weather conditions or stability of the atmosphere.
7.3 Settings for Scanning Conditions
Range of
measurement *1)
Reflectivity
Range mode
Detail
High
Speed
Low
Power
Standard
Close
9% reflection
-
-
-
-
40m
18% reflection
40m
90m
90m
150m
-
90% reflection
100m
210m
210m
350m
-
: Detail
: High Speed
: Low Power
: Standard
: Close
Range mode
Pulse Select
Resolution
(a predetermined value)
Scanning range
Resolution
(a desired value)
Estimated time for scanning