22-1
22 ERROR DISPLAYS
22
ERROR DISPLAYS
Error code
Description
Countermeasures
3 points
required
When area calculating, there are less
than 3 points coordinate data in
selected file.
Confirm the file data and calculate
again.
CALC
ERROR
Calculation is impossible from the
data input.
Confirm the input data.
DELETE
ERROR
When deleting coordinate data, it can not
be done successful.
Confirm the data and delete again.
SIGNAL
OVER
Displayed when you measure a prism of short
distance with non-prism mode, or when the
telescope received too much light.
Measure the distance by using a target
without using a prism, or change the mode
from non-prism mode to prism mode.
E35
Displayed when REM measurement
carried out to the range from zenith or
nadir ±6°.
Operate in the range out of ± 6° from the
zenith or nadir.
E60’s
Any abnormality occurs with EDM (distance
measuring system).
Repair is required.
E71
Displayed when vertical angle 0 position is
set with incorrect procedure.
Confirm the procedure and readjust.
E72
Displayed when Vertical angle -position is
adjusted in wrong position.
Repair is required.
E73
The instrument was not leveled when Vertical
angle 0-position is adjusted.
Level the instrument then carry the
adjustment work.
E80’s
Mainly at the time data transmission
between GPT-3000 series and external
instrument.
Confirm operation procedure is correct or the
connection cables are correct.
E90’s
Abnormality in internal memory
system.
Repair is required.
FILE
EXISTS
The same file name exists.
Use another file name.
FULL
FILES
When making a file, 30 files already exist.
If necessary, send or delete files.
FAILED
INITIALIZE
Initializing can not be done successful.
Confirm initializing data and try to
initialize again.
LIMIT
OVER
Limit of input data exceeded.
Input again.
MEMORY
ERROR
Any abnormality occurs with internal
memory.
Initialize the internal memory.
MEMORY
POOR
Shortage of capacity of the internal
memory.
Download data from internal memory
to PC.
NO DATA
The data is not found in the search
mode
Confirm the data and search again.
NO FILE
There is no file in internal memory.
If necessary, make files.
FILE NOT
SELECTED
When using a file, no file is selected.
Confirm the file and select a file.
P1-P2
distance
too short
When in point to line measurement, the
horizontal distance between first point and
second point is within 1m.
The horizontal distance between first
point and second point must be more
than 1m.