106
FUNCTION SETTING USING SETTINGS SCREEN
COMM (DATA COMMUNICATION)
COMM contains settings related to data input/output with the external device.
Set item display
Setting Change button
Set contents
Initial value
Page
RS-232C
ON
Communication is done by RS-232C.
OFF
OFF
Communication is not done by RS-232C.
Output Data
REF
Only REF data are outputted.
REF/KRT
KRT
Only KRT data are outputted.
R/K
Both REF and KRT data are outputted.
RS Format
STD1
TOPCON STD1 format
STD1
STD2
TOPCON STD2 format
RS Baudrate
2400
TOPCON STD1 format
2400
9600
TOPCON STD2 format
NETWORK (LAN CONNECTION)
NETWORK contains settings related to data input/output in the LAN.
Set item display
Setting Change button
Set contents
Initial value
Page
Data Transfer
Method
TCP/IP Transfer
Data Transfer method is set to "TCP/IP Transfer."
TCP/IP
Transfer
Shared Folder
Transfer
Data Transfer method is set to "Shared Folder
Transfer."
Sever IP Address Set by ten-key display.
IP address of the PC connected for data output is
set.
NONE
Sever Port No.
Set by ten-key display.
Server port No. of the PC connected for data output
is set.
50000
Shared Folder
Setting
Set by keyboard
display
Shared folder is set.
NONE
Print image
output
ON
Print image is outputted when DATA OUT button is
pressed.
ON
OFF
Print image is not outputted when DATA OUT button
is pressed.
KR-1W data
output
ON
When DATA OUT button is pressed, internal printing
is done, and measurement value data are
outputted to the connected PC.
OFF
OFF
When DATA OUT button is pressed, internal printing
is done, but measurement value data are
not outputted to the connected PC.
Screen shot
output
ON
When DATA OUT button is pressed, internal printing
is done, and screen shot are outputted to the
connected PC.
OFF
OFF
When DATA OUT button is pressed, internal printing
is done, but screen shot data are not outputted
to the connected PC.
Ref/Krt data
output (STD2)
ON
When DATA OUT button is pressed, internal printing
is done, and REF/KER data are outputted
to the connected PC.
OFF
OFF
When DATA OUT button is pressed, internal printing
is done, but REF/KER data are not outputted
to the connected PC.
KR-1W
IP address
Set by ten-key display.
IP address of KR-1W is set.
NONE
KR-1W
Subnet Mask
Set by ten-key display.
Subnet mask of KR-1W is set.
NONE
KR-1W
Default Gateway
Set by ten-key display.
Default gateway of KR-1W is set.
NONE
SPECIAL
SPECIAL is the mode for service engineer and only; it is not recommended to be used.
Summary of Contents for KR-1W
Page 1: ...INSTRUCTION MANUAL WAVE FRONT ANALYZER KR 1W ...
Page 2: ......
Page 186: ......