137
FUNCTION SETTING USING SETTINGS
• When the measurement value is more than the threshold value of Corneal Irregular Astig-
matism 2
Threshold value of Corneal Irregular Astigmatism 1: Lowest setting of corneal irregular astig-
matism threshold value
Threshold value of Corneal Irregular Astigmatism 2: Highest setting of corneal irregular astig-
matism threshold value
The threshold value can be set between 0.00 and 100.00 in 0.01 unit.
Before shipment, the default setting is "0.30."
1
On the SETTINGS screen, tap the Set Condition button of the set item display "Corneal
Irregular Astig.1"of "DISPLAY."
2
Enter the value by the ten-key display of the Setting Change button .
3
Setting is complete by pressing the OK button .
SETTING CORNEAL IRREGULAR ASTIG.2 (VALUE 1<VALUE 2)
The Corneal Irregular Astig.2 of the IOL Selection map is set.
The threshold value can be set between 0.00 and 100.00 in 0.01 unit.
Before shipment, the default setting is "0.60."
1
On the SETTINGS screen, tap the Set Condition button of the set item display "Corneal
Irregular Astig.2" of "DISPLAY."
2
Enter the value by the ten-key display of the Setting Change button .
3
Setting is complete by pressing the OK button .
SETTING THE THRESHOLD K VALUE
The threshold K value of the IOL Selection map is set.
• In the following cases, the character color is changed based on the set threshold K value:
• When the K value is within the set value range
• When the K value is outside the set value range
The threshold value can be set between 0.00 and 100.00 in 0.01 unit.
Before shipment, the default setting is "0.50."
1
On the SETTINGS screen, tap the Set Condition button of the set item display "K Read-
ings" of "DISPLAY."
2
Enter the value by the ten-key display of the Setting Change button .
3
Setting is complete by pressing the OK button .